BIRTHNING BED AVE 2 SERVICE MANUAL
Date
06-2017
Version
1.2
6.6.3.4.
Disasembling the Back Section
1. Disassemble the section seat, see Chapter 6.6.2.1.
2. Adjust the height of the lift column of the bed
to be approx. 15 cm below the maximum level of the vertical lift.
3. Lift the back section up to its maximum position.
4. Take away the left plastic cover of the back section and disconnect cables from the
left back-section interface getting it connected to the interface of the upper frame. Viz.
Chapter 6.8.3.1.
5. Disconnect the earthing wire and loosen the bottom bolt of CPR carrier to be able
to disengage the cables. (Picture 6.6.3.4.1.).
6. Pull the cables through below CPR carrier and disengage the cables. (Picture
6.6.3.4.2.).
Picture 6.6.3.4.1.
Picture 6.6.3.4.2.
7. Cut off the tightening strip keeping the actuator cable. Unsecure the securing ring of
the connector and disconnect the back-section actuator. (Picture 6.6.3.4.3.)
Picture 6.6.3.4.3.
Picture 6.6.3.4.4.