Universal Flow Computer UR06
Operating Manual
Page 52 of 55
Subject to changes
A-EN-08728-URRev.B
Bopp & Reuther
Messtechnik GmbH
The application of cables stranded in pairs is recommended. Shielded cables are not required.
However, if shielded cables are used, for example if recommended by the transmitter manufacturer,
the shielding should not be connected to UR06.
Example:
Temperature sensor
4-wires, stranded in pairs
e.g. LiYY (TP) 4*0.5 mm²
Pressure transmitter
2-wires, stranded in pairs
e.g. LiYY (TP) 2*0.5 mm²
DT31x (up to 45 m)
6-wires, stranded, shielded
e.g. LiYCY 6*0.5 mm²
19 Certificates and Approvals
DIN-EN 9001 certified production
Type examination according to EU Directive 2014/32/EU Annex MI-004.
DE-08-MI004-PTB004 (different software version)
Type examination according to German Eichordnung EO 22 as calculator for cold-meters and
combined heat- and cold-meters.
PTB Z 22.75 / 09.02 (different software version)
Type examination according to EU directive 2014/32/EU Annex MI-005 (OIML test report)
CE mark:
The measuring system fulfills the legal requirements of the EC Directives 2014/30/EU (EMC
Directive) and 2014/35/EU (Low Voltage Directive). We confirm successful device testing by
affixing of the CE mark.
20 Appendix
20.1 Error and warning messages
Active errors and warnings are displayed with the error number and a text.
After eliminating the cause of the error, the messages are automatically deleted. Only pulse errors
have to be acknowledged. If all the errors have been eliminated, a message appears stating that new
errors have been listed in the logbook. The last 100 error messages are listed in the logbook with the
time of their occurrence and elimination.
If a measured value is unavailable, the default value (substitute value) is used for further calculations,
but not if the main measured value is affected (dp, flow rate). Flow rate, power and current outputs are
calculated with the default values, the counters are inactive. If fault counters have been configured,
they continue to operate.
Code
Designation
Possible reason
Effect
E01
Power off
No power supply, defective fuse
No functions
Display is black
Message only in the logbook
E02
AD converter I
AD converter for current measuring
defective
Measured value fails or
calculation is carried out with
default values
Error can only be eliminated
by the manufacturer.
E03
AD converter PT
AD converter for temperature
measuring defective
Measured value fails or
calculation is carried out with
default values
Error can only be eliminated
by the manufacturer.
E04
E05
E06
E07
AD-I1
AD-I2
AD-I3
AD-I4
Measured variable outside the
measuring range
Wiring error
Measured value fails or
calculation is carried out with
default values
E08
E09
AD-PT1
AD-PT2
Measured variable outside the
measuring range
Wiring error
Measured value fails or
calculation is carried out with
default values
E10
EEPROM
Error when writing in the internal
EEPROM