Step 18.
Re move the screws that se cure the left-hand pole piece to the coil hous ing and loosen the
screws that se cure that pole piece to the source base plate (see Fig ure 5-8).
Step 19.
Step : Re peat the same pro cess on the right-hand pole piece.
Step 20.
You should now be able to lift the coil hous ing free of the pole pieces, as shown in Fig ure
5-9. If not, fur ther loosen the bot tom row of screws that se cure one of the pole pieces to the
source base plate.
Source Disassembly
STIH-270-2PT
0101-9004-0A
5-9
Figure 5-8. Removing the screws that secure the coil housing to the
pole pieces
Figure 5-7. Uncoupling the water-crossover tube