![BOC Edwards iL600 Instruction Manual Download Page 42](http://html1.mh-extra.com/html/boc-edwards/il600/il600_instruction-manual_2781982042.webp)
Parameter
Unit
Default setpoint values
Low alarm
Low
warning
High
warning
High alarm
LCDP power consumption
kW
ADJ
0.0
3.4
3.8
LCMB power consumption *
kW
- #
- #
- #
- #
* Il600 only.
# LCMB power threshold is 7.3 kW; maximum time in overload is 15 seconds.
Table 2-3 - Electrics and Pump Control Module default setpoints
Parameter
Unit
Default setpoint values
Low alarm
Low
warning
High
warning
High alarm
LCDP pump temperature
o
C
ADJ
ADJ
129
159
LCDP motor cooling-water
return temperature
o
C
ADJ
ADJ
44
ADJ
LCMB pump temperature
*
o
C
ADJ
ADJ
99
119
* iL600 only.
Table 2-4 - Sensor Monitoring Module default setpoints
(sensor-configured iL systems only)
Parameter
Unit
Default setpoint values
Low alarm
Low
warning
High
warning
High alarm
Shaft-seals purge pressure
psig
2.7
3.0
5.0
6.5
Table 2-5 - Gas Module default setpoints
(sensor-configured iL systems only)
iL Dry Pumping Systems
2-13