Warning message
Advisory message
Meaning/recommended action
WARNING
5101
The Sensor Monitoring Module may be
disconnected. If the wiring connections are
correct, the PCA may be faulty.
WARNING
5113
There is a fault in the ADC (analogue-to-
digital convertor) in the Sensor Monitoring
Module; the ADC must be replaced.
WARNING
5411
MB TEMP
The temperature of the LCMB pump is too
high. Check the pumping system inlet
pressure; has the pumping system been
operating with inlet pressure > 5 mbar
(5 x 10
2
Pa) for 1 hour or more. If the
maximum inlet pressure has not been
exceeded, or if the pumping system has not
been operated at the maximum inlet pressure
for longer than the specified time, the
cooling-circuit of the pump maybe blocked.
WARNING
5413
Check that the LCMB pump temperature
sensor is fitted and is correctly connected. If
the sensor is correctly fitted and connected, it
may have failed. Check the operation of the
sensor and replace it if it has failed.
WARNING
5511
DP MOTOR TEMP
MOTOR WATER LOW
CHECK WATER
The temperature of the LCDP pump- motor is
too high. Check the cooling-water connections
in the iL system and check that the
cooling-water supply is on and is as specified
in Section 2. If the connections and supply are
correct, the cooling-circuit of the pump maybe
blocked.
WARNING
5513
The LCMB pump-motor surface temperature
sensor is not fitted, or is disconnected or has
failed.
WARNING
5709
DP TEMP LOW
DP TEMP LOW
SEE MANUAL
The operating temperature of the LCDP pump
is too low.
WARNING
5711
DP TEMP HIGH
DP TEMP HIGH
SEE MANUAL
The operating temperature of the LCDP pump
is too high: there may be process debris in the
pump or the pump may be damaged.
WARNING
5713
The LCDP pump-body temperature sensor is
not fitted, or is disconnected or has failed.
Check and rectify as necessary.
WARNING
6001
WATER FLOW LOW
CHECK COOLING
WATER SUPPLY
The flow of cooling-water through the iL
system is too low. Refer to the recommended
actions for error number 5411.
6-18
iL Dry Pumping Systems
Table 6-1 - Warning messages (continued)