6-19
Gate:
Enables the GATE input for the channel by setting the
method of output control used with the gating
function.
Logic:
Sets the logic level used with the gating function,
either active high or active low.
“Pulse Inhibit” method
T
he gate prevents the channel from being triggered by the channel’s
trigger source pulse. If a pulse has already started when the gate disables
the channel, the pulse will continue normal output but will not restart on
the next trigger pulse.
“Output Inhibit” method
The gate leaves the base triggering alone and enables/ disables the
output directly.
Clocks/Rate Menus (FUNC + RATE key)
Internal Reference Menu
Source:
System Osc
Source:
10MHz Ext
Source:
20MHz Ext
Source:
25MHz Ext
Source:
40MHz Ext
T
0
Period
T
0
Period
T
0
Period
T
0
Period
T
0
Period
Source:
50MHz Ext
Source:
80MHz Ext
Source:
100MHz Ext
T
0
Period
T
0
Period
Setting the Internal Reference Source and Rate
Source:
Selects the internal or external clock source from
which the unit will operate.
To:
Sets the T
0
period which determines the fundamental
output frequency of the unit.