![BNC 575 Series Operating Manual Download Page 17](http://html1.mh-extra.com/html/bnc/575-series/575-series_operating-manual_2781271017.webp)
6-17
Mode Menu
Channel Enable
Channel Enable
Channel Enable
Channel Enable
Mode: Normal
Mode: Single
Shot
Mode: Burst
Mode:
Duty Cycle
#/On Pulses
#/Burst Pulses
#/Off Pulses
Wait Menu
Channel Enable
#/Wait Pulses
Enabling Channel Output
At the top of each of the channel menu pages is a parameter to enable or disable
the channel. Each channel may be individually controlled. When enabled, the
channel key will illuminate.
Setting the Channel Timing Parameters
To define a pulse requires two parameters: the delay to the active edge and the
width of the pulse.
Wid:
Sets the channel pulse width.
Dly:
Sets the channel delay until the active edge.
Setting Pulse Configuration Output Type
The 575 supports two types of outputs: a high speed TTL/CMOS compatible
output and for applications which require different voltage levels or higher
cur-rent, an adjustable voltage output. The pulses can also be defined to be
active high or active low.
Out:
Selects the output mode; TTL/CMOS, Adjustable, Optical,
High Impedance (Hi Z), or Low Impedance (Lo Z)
Pol:
Sets the pulse polarity, active high or active low.
Ampl:
Sets the output voltage level when in the Adjustable mode.
Setting Channel Mode of Operation
Each channel may be set independently to operate in one of four modes: normal,
single shot, burst, or duty cycle (within the CHANNEL menus):
Mode:
Selects the mode for the current channel. Additional
parameters are provided for the burst mode and the duty
cycle mode.
Brst:
Sets the number of pulses in the burst mode to gener-ate
before inhibiting output.