40
Note: The simulated PLD will output at the same frequency as the step frequency.
Recall and Run Program
Once the program is configured and stored into memory, they can be recalled and triggered to
run.
Recall Program
Follow the steps below to recall a stored program from memory:
1.
From the LIST menu (press (Shift)
and press “List” (“8”)), use the arrow keys to select
Recall and press
Enter
Enter
. The display will say RECALL LIST with the prompt Recall List =
0.
2.
This is the location number where the program is stored in. Select a number between 0
to 9 with the rotary knob or the numeric keypad, then press
Enter
Enter
.
3.
Once selected, the display will read “Recall Data Success” and will return back to the
LIST menu.
Run Program
To run a program list after it is recalled, the trigger source must be configured first. Refer to
“Configure List Setting” to set up the trigger source settings.
Follow the steps below to run the program:
1.
Assuming the stored program has already been recalled, from the LIST menu, with
Disable selected, press the up or down key to select Enable and press
Enter
Enter
. The
Trig indicator will appear at the top middle of the display, as shown below:
Note: The “0” next to “List” in the display above is the step number indicator.
2.
If the Trigger Source configured in Configure List Setting is set to Manual (Def), then the
front panel trigger key (Press (Shift)
and then
Enter
Enter
) or the
On/Off
key (see List-
Set section to configure this option) can be used to initiate the program.
300.0V
0.0mA
0.0 0 W
OFF
Trig
PF = 0.000
500
List 0