76
Sweep Time
10 ms to 100 s
Input and Output
Trigger IN
TTL compatible
Maximum rate 1 MHz
Minimum width > 50 ns
Input impedance 1 kΩ
Sync OUT
TTL pulse at programmed frequency
50 Ω source impedance
Modulation IN
5 V
p-p
for 100% modulation
10 kΩ input impedance
DC to > 20 kHz minimum bandwidth
Counter Characteristics
Range
50 Hz to 50 MHz
Resolution
Auto ranging, up to 8 digits
Accuracy
± 0.02 % ± 2 digits
Sensitivity
25 mV
rms
typical
General
Memory Storage
Store up to 20 instrument settings
Arbitrary memory
16,382 points in flash memory
Power Requirements
100 V – 240 V AC ± 10% (90 V – 264 VAC), 47 – 63 Hz
Max. Power Consumption
< 30 VA
Operating Temperature
0 °C – 50 °C
Storage Temperature
-10 °C – 70 °C
Humidity
95% RH, 0 °C – 30 °C
Dimensions
213 mm x 88 mm x 210 mm (WxHxD)
Weight
Approx. 2.5 kg
Safety Standards
EN55011 for radiated and conducted emissions
EN55082
EN61010
*3 V
p-p
into 50 Ω.
Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an
ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change without notice.
www.
GlobalTestSupply
.com
Find Quality Products Online at: