30
3.1.6 PULSE Key
Selects the Pulse waveform menu shown below:
Figure 20 – Pulse Menu
F1: Frequency/Period - Selects the parameter definition of the Pulse repetition period.
F2: Amplitude
- Selects the Amplitude parameter.
In Arbitrary mode, this setting defines the maximum peak-to-peak amplitude of
a full-scale waveform. If the waveform does not use the full scale (data points
from -8191 to +8191), then its actual amplitude will be smaller.
F3: Offset
- Selects the Offset parameter. Change the offset by using the cursor keys,
rotary dial or numerical keys. If a certain setting cannot be produced, the
waveform generator will display a “Setting Conflict” error message.
F4: Pulse
- Selects the Pulse Width and edge edit menu