BT-740 Operation manual
80
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
10.8 Technical description
Accessories have passed the biocompatibility test and meet the requirements of ISO
10993-1.
Fluke's index 2XL Oxygen Analyzer can be used to check the function of the monitor and
can be used to assess the accuracy of the pulse rate but cannot be used to assess the
accuracy of blood oxygen.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry probe and
pulse oximetry.
Measure the maximum temperature between the oxygen probe and the tissue contact
surface
:
Measured as described in Annex BB of ISO 80601-2-61,the temperature is less
than 41 ° C.