BT-740 Operation manual
51
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
6.3.1 Visual alarm
When an alarm occurs, the alarm indicator will flash in different colors and frequencies to
prompt the alarm priority.
High priority alarm: Red, fast flashes.
Medium priority alarm: Yellow, slow flashes.
Low priority alarm: Yellow, lit without flashing.
6.3.2 Audible alarm
An audible alarm is that the monitor prompts the alarm priorities with different sound
characteristics when an alarm occurs.
High priority alarm: Beep-beep-beep--beep-beep----beep-beep-beep--beep-beep
Medium priority alarm: Beep-beep-beep
Low priority alarm: Beep
6.3.3 Alarm information
Alarm information displayed on the physiological or technical alarm area of the monitor
indicates the corresponding alarm information when an alarm occurs. The system will
distinguish the alarm priority with different background colors:
High priority alarm: Red
Medium priority alarm: Yellow
Low priority alarm: Yellow
The following flags in front of physiological alarms are used to distinguish the alarm priorities.
High priority alarm: ***
Medium priority alarm: **
Low priority alarm: *