BT-740 Operation manual
53
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
6.4.2 Alarm sound off
The alarm sound can be turned off through the following operations:
Select
[Settings] smart h
otkey→ “
Settings”.
Select “User Maintenance >>”→enter the password and confirm →“User Maintenance”
menu.
Select “Alarm Param >>”→“ Alarm
Param” menu.
Set “Minimum Alarm Volume” to “0”. “Minimum Alarm Volume” range from 0 to 9, the
default value is 1.
Select
【
Volume
】
smart hotkey →“Volume Setup” menu.
Set “Alarm Volume” to “0”.
When the alarm sound is turned off, the alarm state area on the screen shows the
icon.
If “Minimum Alarm Volume” is larger than 0, the system will cancel alarm sound off state.
WARNING
When the alarm is off, and the alarm reminder signal is on, the system will have
alarm reminder tone.
6.4.3 Alarm pause
Press the
[Pause] smart hotkey or
[Alarm pause] key on the monitor panel to
temporarily stop the alarm of the monitor in the following steps:
[Pause]smart hotkey will appear magnified and reverse colored icon.
The light alarm and audible alarm of the physiological alarms are suspended, and the
alarm information is not displayed.
The remaining time of alarm pause is displayed in the physiological alarm area.
Alarm parameters and upper / lower limit stop flashing.
The audible alarm and light alarm of technical alarms are suspended, but the alarm
message is still displayed.
After the alarm pause is finished, the monitor will automatically cancel the alarm pause state.
During the alarm pause, you can also press
[Pause] smart hotkey or
[Alarm pause] key
on the monitor panel to cancel the alarm pause manually.