BT-740 Operation manual
56
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
6.5.3 Setting a parameter alarm
You can set the parameter alarm for every alarm separately. For SpO
2
, as an example, select
“Alarm Setup >>” in the “Settings” menu and select “SpO
2
” and enter the SpO
2
alarm setup
menu.
1.
Turn on / off alarm
Select “Alarm Switch” and set the alarm switch as follows:
“On”: Turn on SpO
2
alarm; when the parameter alarm occurs, the monitor will
prompt according to the set alarm level.
“Off”: Turn off SpO
2
alarm; icon is displayed in the parameter area, and the
monitor won’t prompt the parameter alarm.
2.
Set the alarm priority
Select “Alarm Level”, and set the alarm priority as follows:
“Mid”: Set the alarm priority to medium.
“High”: Set the alarm priority to high.
Note: Regulatory requirements, the parameter (ECG, blood oxygen saturation, blood pressure)
can set the alarm priority high and mid.
3.
Set the alarm limit
In any cases, the alarm system only allows setting the values within the effective range of the
system, and the upper alarm limit must be higher than the lower alarm limit.
Select “SpO
2
Low Limit” and set the lower limit of SpO
2
alarm.
Select “SpO
2
High Limit” and set the upper limit of SpO
2
alarm.
Select “PR Low Limit” and set the lower limit of PR alarm.
Select “PR High Limit” and set the upper limit of PR alarm.
Type
Adults
Pediatric
Range
Default
Range
Default
SpO2 Low Limit
0-99
90
0-99
90
SpO2 High Limit
1-100
100
1-100
100
PR Low Limit
15-299
50
15-349
75
PR High Limit
16-300
120
16-350
160
4.
Restore default alarm setup