Calibration Procedures
Section 5
46
Calibration Procedures
Sensor replies:
CAL COMPLETE
REMOVE REF STD
Note: Do not remove the calibration reference plaque at this point.
STEP 10.
Wait for the third data message to be received at the PC.
STEP 11.
Note the forward-scatter MOR (located in 4
th
field) in the sensor data
message.
STEP 12.
Send the BB? Command and note the value.
STEP 13.
If the MOR reported is the same as the MOR value of the plaque (±0.01)
and the response to the BB? command is within 5% of the Backscatter
calibration value of the plaque then the sensor is within its calibration
limits. The sensor can be returned to its operational configuration with
confidence. The calibration reference plaque can now be removed.