Section 5
Calibration Procedures
Calibration Procedures
45
STEP 1.
Set up the sensor with the calibration reference plaque in place - see
section 5.1 (power to the sensor need not be removed).
STEP 2.
Send command CO.
Sensor replies: OK.
STEP 3.
Send command: CE.
Sensor replies:
CLEAN WINDOWS,
BLOCK FWD SCAT RCVR OPTICS,
BLOCK TRANSMITTER OPTICS
BLOCK BK SCAT RCVR OPTICS (
not for the SWS-100 – LW)
INSTALL REF STD,
ENTER FWD SCAT VALUE
FORM: XXX.XX
STEP 4.
Fit the grey foam plugs (supplied with the calibration kit) against all three
of the windows (only two needed for the SWS-100 – LW).
STEP 5.
Enter Forward CAL value from the calibration plaque.
STEP 6.
SWS-200 – LW Only
Sensor replies:
ENTER BACK SCAT VALUE
FORM: XXX.XX
Enter Backscatter CAL value from the calibration reference plaque.
STEP 7.
Sensor replies:
CAL IN PROGRESS
Wait for approximately 2 minutes.
Sensor replies:
REMOVE OPTICS BLOCKS,
ENTER "OK"
STEP 8.
Remove grey foam plugs from all windows and send text: OK.
Sensor replies:
CAL CONTINUES
STEP 9.
Wait for approximately 2 minutes.