Sensor Set-up
Section 1
14
STEP 4 – Configuration Options
1.4.7
Baud Rate Configuration
Default communication parameters are 9600 Baud, 8 data bit, 1 stop bit, no parity,
and no flow control. The baud rate may be changed if required as follows.
Send
%B(Number)
Just typing %B will bring up the different baud rate options:
SELECT REQUIRED BAUDRATE BY TYPING %B(NUMBER)
1....1200 BAUD
2....2400 BAUD
3....4800 BAUD
4....9600 BAUD
5....19K2 BAUD
6....38K4 BAUD
7....57K6 BAUD
Select the baud rate to use, for example to select 4800 baud the user would type
%B3<CRLF>
The user then receives a prompt to send an "OK" to the sensor at the new baud rate
setting. The new setting will only be accepted if the user manages to communicate
with the sensor at the new baud rate within 60 seconds. Otherwise the sensor will
reset and continue operation with the original baud rate settings. If an "OK"
command is received at the new baud rate the sensor will update its settings and
restart.