![BIOTRONIK ICD Series Technical Manual Download Page 46](http://html1.mh-extra.com/html/biotronik/icd-series/icd-series_technical-manual_2756135046.webp)
44
Technical Data
Ilesto 7 DR-T
Service times with GB 2992 battery:
Ilesto 7 HF-T
Service times with GB 2992 battery:
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
10.73
8.82
7.48
6.50
5.74
15%
10.22
8.47
7.23
6.31
5.59
50%
9.20
7.76
6.70
5.90
5.27
100%
8.05
6.92
6.07
5.40
4.87
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
9.96
8.29
7.10
6.20
5.51
15%
9.33
7.85
6.77
5.95
5.31
50%
8.12
6.97
6.11
5.43
4.89
100%
6.85
6.01
5.36
4.83
4.40