Parameters
4
Parameters
Parameters4GA-HW_en--mul_393468-BTechnical[nbsp ]Manual for the[nbsp ]DeviceIlesto 5/7 VR-T, VR-T DX, DR-T, HF-T
Bradycardia / CRT
General ICD therapy
Timing: Basic rate day/
night and rate hystereses
Parameter
Range of values
Standard
VR
DX
DR
HF
ICD therapy
OFF; ON
ON
x
x
x
x
Programs
Display standard program; Display
safe program; Display first interro-
gated program; Individual 1,2,3
–
x
x
x
x
Parameter
Range of values
Standard
VR
DX
DR
HF
Basic rate
30 ... (5) ... 100 ... (10) ... 160 bpm
40 bpm
x
x
60 bpm
x
x
Night rate
OFF; 30 ... (5) ... 100 bpm
OFF
x
x
x
x
Begin of night
00:00 ... (00:01) ... 23:59 hh:mm
06:00 hh:mm
x
x
x
x
End of night
22:00 hh:mm
Rate hysteresis
OFF
-5 ... (-5) ... -25 ... (-20) ... -65 bpm
OFF
x
x
x
x
Scan/repetitive
OFF; ON
ON
x
x
x
x