33
Parameters
Tachycardia
Detection
Therapy: ATP
Parameter
Range of values
Standard
VR
DX
DR
HF
Interval AT/AF
240 ... 600 ms
300 ms
x
x
x
Interval VT1
OFF; 270 ... (10) ... 600 ms
OFF
x
x
x
x
Interval VT2
OFF; 270 ... (10) ... 500 ms
Interval VF
OFF; 240 ... (10) ... 400 ms
300 ms
Detection counter VT1
10 ... (2) ... 60
26
x
x
x
x
Detection counter VT2
10 ... (2) ... 40
16
Detection counter VF
6 out of 8; 8 out of 12; 10 out of 12;
12 out of 16; 14 out of 16; 16 out of 20;
18 out of 24; 20 out of 24; 20 out of 24;
22 out of 24; 24 out of 30; 28 out of 30
8 out of 12
Redetection counter VT1
10 ... (2) ... 30
20
x
x
x
x
Redetection counter VT2
14
SMART detection VT1/VT2
OFF; ON
ON
x
x
x
SMART detection ON:
– Onset VT1/VT2
4 ... (4) ... 32%
20%
x
x
x
– Stability VT1/VT2
8 ... (4) ... 48%
12%
SMART detection OFF:
– Onset VT1/VT2
OFF; 4 ... (4) ... 32%
20%
x
x
x
x
– Stability VT1/VT2
OFF; 8 ... (4) ... 48 ms
24 ms
Sustained VT
OFF; 1; 2; 3; 5; 10; 20; 30 min
OFF
x
x
x
x
Forced termination
OFF; 1 ... (1) ... 10 min
1 min
x
x
x
Parameter
Range of values
Standard
VR
DX
DR
HF
ATP type for VT1/VT2
Burst; Ramp
OFF
x
x
x
x
ATP type for VF
OFF; Burst; Ramp
Burst
x
x
x
x
ATP optimization
OFF; ON
OFF
x
x
x
x
Attempts
OFF; 1 ... (1) ... 10
OFF
x
x
x
x
Number S1 for VT1/VT2
1 ... (1) ... 10
5
x
x
x
x
Number S1 for VF
8
S1 decrement for VT1/VT2 and for VF
5 ... (5) ... 40 ms
10 ms
x
x
x
x
Scan decrement
OFF; 5 ... (5) ... 40 ms
OFF
x
x
x
x
Additional S1 for VT1/VT2
OFF; ON
ON
x
x
x
x
Ventricular pacing for VT1/VT2
RV; LV; BiV
RV
x
Ventricular pacing for VF
RV
R-S1 interval for VT1/VT2
70 ... (5) ... 95%
80%
x
x
x
x
R-S1 interval for VF
85%