![BIOTRONIK ICD Series Technical Manual Download Page 44](http://html1.mh-extra.com/html/biotronik/icd-series/icd-series_technical-manual_2756135044.webp)
42
Technical Data
Battery Data
Battery characteristics
The following data is provided by the manufacturers:
Storage period
The storage period affects the battery service time.
•
Devices should be implanted within 19 months between the date of manufac-
ture and the use by date (indicated on the package).
•
If the ICD is implanted shortly before the use by date, the expected service time
may be reduced by up to 16 months.
Calculation of service times
•
The services times have been calculated as follows – in all chambers
depending on the device type:
— Pulse amplitude: 2.5 V
— Pulse width: 0.4 ms
— Pacing impedance: 500 Ω
— Basic rate: 60 bpm
— Home Monitoring: ON, 1 device message each day and 12 transmissions of
an IEGM online HD per year
— Diagnostic functions and recordings: permanently set
•
Capacitor reforming is performed 4 times per year and therefore at least
4 maximum charges for shocks have to be assumed per year even if less than 4
are delivered.
Calculation of the number
of shocks
Calculation of the number of shocks: Longevity [in years] x number of shocks per
year
Ilesto 5 VR-T
Service times with GB 2992 or LiS 3410 RA battery:
Manufacturer
GREATBATCH, INC.
Clarence, NY 14031
LITRONIK GmbH & Co
01796 Pirna,
Germany
Battery type
GB 2992
LiS 3410 RA
Battery ID number shown on
the programmer
3
4
Device type
VR, (DX), DR, HF
Battery voltage at ERI
2.5 V
2.85 V
Charge time at BOS
8 s
8 s
Charge time at ERI
10 s
10 s
Usable capacity until ERI
Ilesto 5: 1390 mAh
Ilesto 7: 1600 mAh
1390 mAh
Usable capacity until EOS
1730 mAh
1520 mAh
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
10.42
8.39
7.01
6.03
5.28
15%
10.14
8.20
6.89
5.93
5.21
50%
9.55
7.81
6.60
5.72
5.05
100%
8.81
7.31
6.24
5.45
4.83