36
Parameters
Diagnostics
The following can be set:
The following can additionally be set for Ilesto 7:
Parameter
Range of values
VR
DX
DR
HF
Standard
For AT/AF
OFF; ON
For Ilesto 7: Extended ON
x
x
x
ON
For SVT
OFF; ON
x
x
x
ON
Periodic recording
When Home Monitoring OFF:
OFF; 30 ... (30) ... 180 days
x
x
x
x
90 days
IEGM configuration
RA, RV, LV
RA, RV, FF
FF; RV; LV
x
RA, RV, LV
Parameter
Range of values
VR
DX
DR
HF
Standard
Start resting period
0:00 ... (1:00) ... 23:00 hh:mm
x
x
x
x
2:00 hh:mm
Duration of resting period
0.5 ... (0.5) ... 12 h
x
x
x
x
4 h
AV delay modification in sensing test
OFF; 300 ms
x
x
x
300 ms
Thoracic impedance (TI)
OFF, ON
x
x
x
x
OFF