General Safety Instructions
2
General Safety Instructions
General Safety Instructions2GA-HW_en--mul_393468-BTechnical[nbsp ]Manual for the[nbsp ]DeviceIlesto 5/7 VR-T, VR-T DX, DR-T, HF-T
Operating Conditions
Care during shipping and
storage
•
Devices are not to be stored or transported close to magnets or sources of
electromagnetic interference.
•
Note the effects of the storage duration; see Battery Data.
Delivery in shipment mode
The device is delivered in shipment mode to protect the battery; capacitor
reforming required during storage could result in controlled extended charge
times of the shock capacitors.
•
The shipment mode is displayed on the programmer after loading the device
program (it is deactivated during implantation on initial measurement of the
pacing impedance).
Temperature
Extremely low and high temperatures affect the service time of the battery in the
device.
•
Temperatures of 5°C to 45°C are permitted for transport, storage, and use.
Sterile delivery
The device and the screwdriver have been gas-sterilized. Sterility is guaranteed
only if the blister and quality control seal have not been damaged.
Sterile container
The device and screwdriver are packaged in two separately sealed blisters. The
inner blister is also sterile on the outside so that it can be transferred in a sterile
state during implantation.
Single use only
The device and screwdriver are intended for single use only.
•
Do not use the device if the package is damaged.
•
The device must not be resterilized and reused.