43
Technical Data
Ilesto 5 VR-T DX
Service times with GB 2992 or LiS 3410 RA battery:
Ilesto 5 DR-T
Service times with GB 2992 or LiS 3410 RA battery:
Ilesto 5 HF-T
Service times with GB 2992 or LiS 3410 RA battery:
Ilesto 7 VR-T
Service times with GB 2992 battery:
Ilesto 7 VR-T DX
Service times with GB 2992 battery:
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
9.48
7.76
6.57
5.70
5.03
15%
9.24
7.61
6.46
5.61
4.96
50%
8.75
7.26
6.21
5.42
4.81
100%
8.12
6.83
5.89
5.17
4.62
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
9.48
7.76
6.57
5.70
5.03
15%
9.02
7.45
6.35
5.53
4.89
50%
8.10
6.81
5.88
5.17
4.61
100%
7.08
6.07
5.32
4.73
4.26
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
8.78
7.29
6.23
5.44
4.82
15%
8.21
6.89
5.94
5.21
4.65
50%
7.14
6.12
5.35
4.76
4.28
100%
6.01
5.27
4.69
4.23
3.85
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
11.78
9.52
7.98
6.87
6.03
15%
11.48
9.32
7.84
6.76
5.95
50%
10.81
8.87
7.52
6.53
5.76
100%
9.99
8.31
7.11
6.21
5.52
Stimulation
Longevity [in years] at number of shocks per year
4
8
12
16
20
0%
10.73
8.82
7.48
6.50
5.74
15%
10.48
8.65
7.36
6.40
5.66
50%
9.92
8.26
7.08
6.19
5.50
100%
9.22
7.77
6.71
5.91
5.27