Biotage ELSD-1080
5.2
Detector Errors
Error Description
Suggestions
10
Air temp <= 10 °C inside
enclosure
Environmental temp too cold.
Move to warmer location
11
Air temp > 40 °C temp inside
enclosure
Environmental temp too hot.
Move to cooler location
12
On-board sensor heater failed
Vapor sensor needs replacing
13
Rear panel sensor heater failed
Check vapor sensor on rear of
ELSD is not damaged/ Vapor
sensor needs replacing
14
Leak detected inside unit
Stop pump and investigate
15
Liquid leak sensor has detected
liquid in drip tray
Stop pump and investigate
16
Fan driver IC thermal shutdown
Fault with the circuitry / wires /
fan Call service engineer if this
happens regularly
17
Fan stopped
Fault with the circuitry / wires /
fan Call service engineer if this
happens regularly
18
Nebulizer temperature exceeded
threshold after stabilizing
Faulty thermocouple or heater
control
19
Evaporator temperature
exceeded threshold after
stabilizing
Faulty thermocouple or heater
control
20
Light source error
Replace light source assembly
21
Evaporator gas flow rate
exceeded threshold after
stabilizing
Check operating head pressure,
otherwise faulty mass flow
controller
22
Invalid nebulizer temp reading
(FAULTY RTD)
Replace nebulizer heater
assembly
23
Invalid evaporator temp reading
(FAULTY RTD)
Replace evaporator assembly
24
Fan failure on cooled evaporator
Fault with the peltier unit. Call
service engineer if this happens
regularly
25
Evaporator tube current outside
of normal range
Fault with the peltier unit. Call
service engineer if this happens
regularly
Table 6. Detector Errors
50
Publication date: July 2010
Summary of Contents for ELSD-1080
Page 4: ...Biotage ELSD 1080 This page is intentionally left blank 4 Publication date July 2010 ...
Page 8: ...Biotage ELSD 1080 This page is intentionally left blank 8 Publication date July 2010 ...
Page 44: ...Biotage ELSD 1080 This page is intentionally left blank 44 Publication date July 2010 ...
Page 54: ...Biotage ELSD 1080 This page is intentionally left blank 54 Publication date July 2010 ...
Page 56: ...Biotage ELSD 1080 This page is intentionally left blank 56 Publication date July 2010 ...
Page 58: ...Biotage ELSD 1080 This page is intentionally left blank 58 Publication date July 2010 ...