Biotage ELSD-1080
1.6
Site Preparation Checklist
Environmental
conditions
Temperature 10 to 35 °C (50 to 95 °F).
At constant temperature.
Avoid positioning in direct sunlight.
Humidity 10-80%.
Power
USA & Japan 115V (AC) ±10%.
50/60 Hz, 2A max, with protective earth
connection.
Europe 230V (AC) ±10%.
50/60 Hz, 2A max, with protective earth
connection.
Note:
Ensure the power switch and appliance
coupler remain accessible at all times.
Bench space
The standard detector will require the following
bench space:
•
width 25 cm (9.8”)
•
depth 60 cm (23.6”)
•
height 41.5 cm (16.3”)
Note:
The detector can be computer controlled
and additional bench space will be required for
the computer.
Gas Supply
Nitrogen (>98% purity, filtered to 0.2 µm).
Air can only be used for non-flammable solvents.
The detector is not calibrated for use with gases
other than air or nitrogen.
Nitrogen generator strongly recommended for
best performance.
For operation with other inert gases contact
Biotage for advice.
Extraction
requirements
The exhaust gases from the detector must be
extracted to a fume hood or similar extraction unit.
The PVC extraction tube provided can be
extended if necessary, provided the diameter of
the extension tube used is >50 mm.
Table 2. Site Preparation Checklist
Warning
Biotage ELSD-1080 detectors weigh approximately
13 kg (28.7lb). To avoid injury to personnel or damage
to equipment, always follow all precautions for lifting
heavy objects when moving the detector. It is advisable
to locate the system on a sturdy bench.
18
Publication date: July 2010
Summary of Contents for ELSD-1080
Page 4: ...Biotage ELSD 1080 This page is intentionally left blank 4 Publication date July 2010 ...
Page 8: ...Biotage ELSD 1080 This page is intentionally left blank 8 Publication date July 2010 ...
Page 44: ...Biotage ELSD 1080 This page is intentionally left blank 44 Publication date July 2010 ...
Page 54: ...Biotage ELSD 1080 This page is intentionally left blank 54 Publication date July 2010 ...
Page 56: ...Biotage ELSD 1080 This page is intentionally left blank 56 Publication date July 2010 ...
Page 58: ...Biotage ELSD 1080 This page is intentionally left blank 58 Publication date July 2010 ...