Do  not  plug  the  unit  into  an  ungrounded  mains  outlet,  and  do  not  use  an 
ungrounded extension lead.  

 

Ensure  that  the  power  switch  and  power  connector  are  easily  accessible  during 
use. 

 

Disconnect  the  unit  from  the  electric  circuit  before  moving  or  performing  servicing 
tasks. 

 

Turn off the unit by disconnecting the mains plug from mains outlet. 

 

If  liquid  penetrates  into  the  unit,  disconnect  it  from  the  external  power  supply  unit 
and have it checked by a repair and maintenance technician. 

 

Do  not  operate  the  unit  in  premises  where  condensation  can  form.  Operating 
conditions of the unit are defined in the Specifications section. 

DURING OPERATION 

 

Do  not  operate  the  unit  in  environments  with  aggressive  or  explosive  chemical 
mixtures.  Please  contact  manufacturer  for  possibility  of  operation  of  the  unit  in 
specific atmosphere. 

 

Do not operate the unit if it is faulty or has been installed incorrectly. 

 

Do not use outside laboratory rooms. 

BIOLOGICAL SAFETY 

 

It is the user’s responsibility to carry out appropriate decontamination if hazardous 
material is split on or penetrates into the equipment. 

2.  General information 

Water  purification  system  “Labaqua”  produces  pure  and  ultrapure  water  that 

complies with ISO 3696 Grade II and Grade I water requirements respectively. 

 

Pure (ISO 3696 Grade II) water applications include, but are not limited to: 

 

Feed  for  laboratory  equipment  (washing  machines,  clinical  analysers,  humidifiers, 
autoclaves, hydrogen gas generators); 

 

Manufacturing of chemical and biochemical reagents; 

 

Buffer preparation; 

 

Microbiological media preparation; 

 

Sensitive analytical techniques (e.g. atomic absorption, ICP-OES). 

 

Ultrapure (ISO 3696 Grade I) water applications include, but are not limited to: 

 

High sensitivity analytical techniques (ICP-MS);  

 

High performance liquid chromatography; 

 

Molecular biology; 

 

Cell culture. 

 

 

Summary of Contents for Labaqua

Page 1: ......

Page 2: ...2...

Page 3: ...4 2 General information 5 3 Getting started 8 4 Rinsing the system 11 5 Operation 12 6 TOC level reduction and monitoring 14 7 Specifications 15 8 Cleaning and maintenance 17 9 Warranty and Claims 18...

Page 4: ...ign of the unit Do not block any ventilation openings Do not operate the unit with its covers removed Do not drink deionized water WATER LEAKAGE SAFETY Make sure that all water connection tubes are un...

Page 5: ...cific atmosphere Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriat...

Page 6: ...omatically switch on the boost pump Fig 1 4 and LowPres message will be shown on the display The boost pump Fig 1 4 is used to maintain pressure at the level necessary for the efficient operation of t...

Page 7: ...s and Grade I water is dispensed via the 0 22 m point of use microfilter Labaqua Bio system has an ultrafilter instead of the microfilter To dispense ISO 3696 Grade II water press the Pump Grade II ke...

Page 8: ...ank with pump level switch and tap 1 set 3 8 with elbow 3m overflow tube 1 pce 3 8 1m tank dispense tube 1 pce tank cable 1 pce hex key for controller module 1 pce Operation manual declaration of conf...

Page 9: ...the new pre filter under warranty only if we get a photo of the sediment filter installed in the tap water line Feed water hardness does not affect produced water quality However operation of the syst...

Page 10: ...move the packing material inside 3 4 8 Connect 1 4 John Guest outlet OUT to the inlet In of the tank 3 4 9 Connect 1 4 John Guest outlet Rec Out to the inlet Rec In of the tank 3 4 10 Connect 1 4 John...

Page 11: ...lled site push the tubes into John Guest fittings at the inner wall Then attach the module fully to the inner wall by putting the latch at the bottom of the module into slot on the wall 3 6 3 Screw th...

Page 12: ...next morning and dispense water in the tank through point of use filter microfilter on Labaqua HPLC or ultrafilter on Labaqua Bio The microfilter or ultrafilter needs to be rinsed with at least 20 li...

Page 13: ...is neither full nor disconnected and the feed water pressure is enough the display shows Running Now the Labaqua system is filling the tank with Grade II water As soon as the tank is full the system...

Page 14: ...le duration time 6 TOC level reduction and monitoring Total organic carbon TOC describes level of organic contaminants in pure water High concentration of organic contaminants may affect results of se...

Page 15: ...lation 7 Specifications The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 15 C to 30 C with maximum change at less than 2 C in 24 hour...

Page 16: ...e pump 50 L BS 070102 DK Water storage tank Comfort with dispense pump 100 L BS 070102 FK Water storage tank Comfort with dispense pump 200 L BS 070102 GK Water storage tank Comfort with dispense pump...

Page 17: ...key Fig 4 3 two times to select and confirm Switch off the system Wait 5 seconds and switch on the system again 8 5 Expendable parts and the replacement interval are shown below Description Replacemen...

Page 18: ...chnical support section on our website at the link below 9 3 Warranty covers only the units transported in the original package 9 4 If any manufacturing defects are discovered by the Customer an unsat...

Page 19: ...19 10 Declaration of Conformity...

Page 20: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 06 April 2017...

Reviews: