3.4.
Setting
up
3.4.1. Install the controller module and connect the cable (Fig. 2/1).
3.4.2. Carefully attach the controller module to the holder at the top right of the system
(Fig. 2/2)
3.4.3. While the system is disconnected from mains, it is recommended to carefully
remove the rear panel and check for any transportation damage. Reinstall the panel
before connecting the power cable to 230V AC power supply socket.
3.4.4. Fix the controller module with two screws on the lower side of the module.
3.4.5. Unpack the microfilter/ultrafilter and screw it into the threaded outlet under the
controller module holder (Fig. 2/3). Before connecting, make sure that the sealing
O-ring is inside the holder.
3.4.6. Connect the feed water to the
Ø1/4’’ John Guest inlet marked
Tap Water
. Use an
appropriate adapter from tap water to
Ø1/4’’ tube, see p. 3.2. Tap water should be
filtered with 5-micron mechanical filter (not included).
3.4.7. Connect
Ø1/4’’ JG connector marked
Drain
to the drain using
Ø1/4’’ plastic tube.
Note:
Unscrew the cap of the purified water storage tank and remove the
packing material inside.
3.4.8. Connect
Ø1/4’’ John Guest outlet “
OUT
”
to the inlet
“
In
”
of the tank.
3.4.9. Connect
Ø1/4’’ John Guest outlet “
Rec Out
”
to the inlet
“
Rec In
”
of the tank.
3.4.10. Connect
Ø1/4’’ John Guest outlet “
Rec In
”
to the inlet
“
Rec Out
”
of the tank.
3.4.11. Connect the tank cable to the
TANK
connector.
Note:
If nothing is connected to the
TANK
connector, the system will show
“TnkFull”.
3.4.12. Connect the power cable. Make sure the system is properly grounded.
3.4.13. Connect approximately 1 m of
Ø3/8’’ tube to the tank
Out
connector to ensure
convenient Grade II water dispensing. The end of the tube should be free and must
not be connected to anything. The backpressure will damage the pump.
3.4.14. Connect the
Overflow
connector of the tank to the drain with
Ø3/8’’ tube.
Fig. 2. Setting up
3.5.
Feed water
and drain requirements
3.5.1. Feed
water should be filtered with 5 μm sediment filter. If the filter is not installed,
the input strainer may become clogged and block the water flow. Life of pre-filters
can be reduced too.
3.5.2. Feed water connector port can be
Ø1/2’’ (male or female) or Ø3/4’’ NPTF (male)
thread. The system is equipped with adapters for both water supply connectors.
The feed water tube should be connected to the