Biosan Labaqua Operating Manual Download Page 10

10 

3.4. 

Setting

 

up

 

3.4.1.  Install the controller module and connect the cable (Fig. 2/1). 
3.4.2.  Carefully  attach  the  controller  module  to  the  holder  at  the  top  right  of  the  system 

(Fig. 2/2) 

3.4.3.  While  the  system  is  disconnected  from  mains,  it  is  recommended  to  carefully 

remove the rear panel and check for any transportation damage. Reinstall the panel 
before connecting the power cable to 230V AC power supply socket. 

3.4.4.  Fix the controller module with two screws on the lower side of the module. 
3.4.5.  Unpack  the  microfilter/ultrafilter  and  screw  it  into  the  threaded  outlet  under  the 

controller  module  holder  (Fig. 2/3).  Before  connecting,  make  sure  that  the  sealing 
O-ring is inside the holder. 

3.4.6.  Connect the feed water to the 

Ø1/4’’ John Guest inlet marked 

Tap Water

. Use an 

appropriate adapter from tap water to 

Ø1/4’’ tube, see p. 3.2. Tap water should be 

filtered with 5-micron mechanical filter (not included). 

3.4.7.  Connect 

Ø1/4’’ JG connector marked 

Drain

 to the drain using 

Ø1/4’’ plastic tube. 

 

Note:

 

Unscrew the cap of the purified water storage tank  and remove the 
packing material inside. 

3.4.8.  Connect 

Ø1/4’’ John Guest outlet “

OUT

 to the inlet 

In

 of the tank. 

3.4.9.  Connect 

Ø1/4’’ John Guest outlet “

Rec Out

 to the inlet 

Rec In

 of the tank. 

3.4.10. Connect 

Ø1/4’’ John Guest outlet “

Rec In

 to the inlet 

Rec Out

 of the tank. 

3.4.11. Connect the tank cable to the 

TANK

 connector.  

 

Note:

 

If nothing is connected to the 

TANK

 connector, the system will show 

“TnkFull”. 

3.4.12. Connect the power cable. Make sure the system is properly grounded. 
3.4.13. Connect  approximately  1  m  of 

Ø3/8’’  tube  to  the  tank 

Out

  connector  to  ensure 

convenient Grade II water dispensing. The end of the tube should be free and must 
not be connected to anything. The backpressure will damage the pump. 

3.4.14. Connect the 

Overflow

 connector of the tank to the drain with 

Ø3/8’’ tube. 

 

 

 

Fig. 2. Setting up 

3.5. 

Feed water

 

and drain requirements

 

3.5.1.  Feed 

water should be filtered with 5 μm sediment filter. If the filter is not installed, 

the  input  strainer  may  become  clogged  and  block  the  water  flow.  Life  of  pre-filters 
can be reduced too. 

3.5.2.  Feed  water  connector  port  can  be 

Ø1/2’’  (male  or  female)  or  Ø3/4’’  NPTF  (male) 

thread.  The  system  is  equipped  with  adapters  for  both  water  supply  connectors. 
The feed water tube should be connected to the 

Ø1/4’’ JG port of the adapter. 

Summary of Contents for Labaqua

Page 1: ......

Page 2: ...2...

Page 3: ...4 2 General information 5 3 Getting started 8 4 Rinsing the system 11 5 Operation 12 6 TOC level reduction and monitoring 14 7 Specifications 15 8 Cleaning and maintenance 17 9 Warranty and Claims 18...

Page 4: ...ign of the unit Do not block any ventilation openings Do not operate the unit with its covers removed Do not drink deionized water WATER LEAKAGE SAFETY Make sure that all water connection tubes are un...

Page 5: ...cific atmosphere Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriat...

Page 6: ...omatically switch on the boost pump Fig 1 4 and LowPres message will be shown on the display The boost pump Fig 1 4 is used to maintain pressure at the level necessary for the efficient operation of t...

Page 7: ...s and Grade I water is dispensed via the 0 22 m point of use microfilter Labaqua Bio system has an ultrafilter instead of the microfilter To dispense ISO 3696 Grade II water press the Pump Grade II ke...

Page 8: ...ank with pump level switch and tap 1 set 3 8 with elbow 3m overflow tube 1 pce 3 8 1m tank dispense tube 1 pce tank cable 1 pce hex key for controller module 1 pce Operation manual declaration of conf...

Page 9: ...the new pre filter under warranty only if we get a photo of the sediment filter installed in the tap water line Feed water hardness does not affect produced water quality However operation of the syst...

Page 10: ...move the packing material inside 3 4 8 Connect 1 4 John Guest outlet OUT to the inlet In of the tank 3 4 9 Connect 1 4 John Guest outlet Rec Out to the inlet Rec In of the tank 3 4 10 Connect 1 4 John...

Page 11: ...lled site push the tubes into John Guest fittings at the inner wall Then attach the module fully to the inner wall by putting the latch at the bottom of the module into slot on the wall 3 6 3 Screw th...

Page 12: ...next morning and dispense water in the tank through point of use filter microfilter on Labaqua HPLC or ultrafilter on Labaqua Bio The microfilter or ultrafilter needs to be rinsed with at least 20 li...

Page 13: ...is neither full nor disconnected and the feed water pressure is enough the display shows Running Now the Labaqua system is filling the tank with Grade II water As soon as the tank is full the system...

Page 14: ...le duration time 6 TOC level reduction and monitoring Total organic carbon TOC describes level of organic contaminants in pure water High concentration of organic contaminants may affect results of se...

Page 15: ...lation 7 Specifications The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 15 C to 30 C with maximum change at less than 2 C in 24 hour...

Page 16: ...e pump 50 L BS 070102 DK Water storage tank Comfort with dispense pump 100 L BS 070102 FK Water storage tank Comfort with dispense pump 200 L BS 070102 GK Water storage tank Comfort with dispense pump...

Page 17: ...key Fig 4 3 two times to select and confirm Switch off the system Wait 5 seconds and switch on the system again 8 5 Expendable parts and the replacement interval are shown below Description Replacemen...

Page 18: ...chnical support section on our website at the link below 9 3 Warranty covers only the units transported in the original package 9 4 If any manufacturing defects are discovered by the Customer an unsat...

Page 19: ...19 10 Declaration of Conformity...

Page 20: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 06 April 2017...

Reviews: