Error messages 

Description 

Filter 

Pre-filter set (P/N 10019) needs replacement 

DI Err. 

Deionization module (P/N 10310) needs replacement 

TankS 

Tank sensor is damaged 

Beeping sound during 
“Running” stage 

Sterilization UV bulb needs replacement. 

 

indication of current water grade (superscript) and water conductivity (precise 
readings of water quality are shown when the system is in Running, Grade I 
dispense or Recirculation mode); 

water temperature and filter life counter. 

5.3. 

If the system is equipped with TOC (total organic carbon) level monitoring module, 
display  shows  TOC  reading  (e.  g.  <2  ppb).  TOC  level  reading  will  be  shown  only 
after the storage tank is at least partially filled with water and recirculation stage has 
started. 

5.4. 

The  purified  water  quality  is  controlled  by  two  sensors.  First  sensor  (Fig. 1/11) 
measures and displays conductivity of Grade II water, flowing into or recirculating in 
the storage tank. Second sensor (Fig. 1/19) measures and displays conductivity of 
Grade I water that is dispensed at the moment. 

5.5. 

Indication in superscript next to the conductivity can take values “1” or “2”, showing 
current water grade. 

5.6. 

To  fill  the  tank  with  purified  water,  press  and  hold  the 

Run

  key  (Fig.  4/8)  for  2 

seconds.  If  tank  is  neither  full  nor  disconnected  and  the  feed  water  pressure  is 
enough,  the  display  shows  “Running”.  Now  the  Labaqua system is  filling  the  tank 
with  Grade  II  water.  As  soon  as  the  tank  is  full,  the  system  shuts  down  the  water 
supply and shows the “TnkFull” message. 

5.7. 

To  dispense  Grade  I  water,  press 

Pump  Grade  I

  key  (Fig.  4/6).  After  five  to  ten 

seconds,  Grade  I  water  is  dispensed  through  the  0.22 

μm  filter.  If  ultrafilter  is 

installed, loosen the degassing valve to get rid of air. Press the 

Pump Grade I

 key 

again to stop water dispensing. 

5.8. 

To dispense Grade II water from the tank, press the 

Pump Grade II

 key (Fig. 4/7). 

To stop dispensing, press the 

Pump Grade II

 key again. 

5.9. 

System will circulate water on periodical basis to prevent growth of microorganisms 
in  the  tank.  During  recirculation  phase,  the  reading  of  Grade  II  water  is  shown  on 
the  display.  For  the  systems  with  TOC  monitoring  module,  display  shows  TOC 
reading in ppb. TOC reading appears in the left bottom display when purified water 
storage tank is full and recirculation stage starts. 

5.10.  Recirculation is switched  off when  the  system  is in  the  Standby  mode.  To  prolong 

life of polishing module, please set the system to the Standby mode if it is not used 
for long time (e.g. for the night).  

5.11.  Press 

Run

 key to toggle Standby mode on and off. 

5.12.  If feed water pressure falls, the 

display indicates “LowPres”. 

 

Note:

 

To actuate a key, press and hold it for two seconds. 

Summary of Contents for Labaqua

Page 1: ......

Page 2: ...2...

Page 3: ...4 2 General information 5 3 Getting started 8 4 Rinsing the system 11 5 Operation 12 6 TOC level reduction and monitoring 14 7 Specifications 15 8 Cleaning and maintenance 17 9 Warranty and Claims 18...

Page 4: ...ign of the unit Do not block any ventilation openings Do not operate the unit with its covers removed Do not drink deionized water WATER LEAKAGE SAFETY Make sure that all water connection tubes are un...

Page 5: ...cific atmosphere Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriat...

Page 6: ...omatically switch on the boost pump Fig 1 4 and LowPres message will be shown on the display The boost pump Fig 1 4 is used to maintain pressure at the level necessary for the efficient operation of t...

Page 7: ...s and Grade I water is dispensed via the 0 22 m point of use microfilter Labaqua Bio system has an ultrafilter instead of the microfilter To dispense ISO 3696 Grade II water press the Pump Grade II ke...

Page 8: ...ank with pump level switch and tap 1 set 3 8 with elbow 3m overflow tube 1 pce 3 8 1m tank dispense tube 1 pce tank cable 1 pce hex key for controller module 1 pce Operation manual declaration of conf...

Page 9: ...the new pre filter under warranty only if we get a photo of the sediment filter installed in the tap water line Feed water hardness does not affect produced water quality However operation of the syst...

Page 10: ...move the packing material inside 3 4 8 Connect 1 4 John Guest outlet OUT to the inlet In of the tank 3 4 9 Connect 1 4 John Guest outlet Rec Out to the inlet Rec In of the tank 3 4 10 Connect 1 4 John...

Page 11: ...lled site push the tubes into John Guest fittings at the inner wall Then attach the module fully to the inner wall by putting the latch at the bottom of the module into slot on the wall 3 6 3 Screw th...

Page 12: ...next morning and dispense water in the tank through point of use filter microfilter on Labaqua HPLC or ultrafilter on Labaqua Bio The microfilter or ultrafilter needs to be rinsed with at least 20 li...

Page 13: ...is neither full nor disconnected and the feed water pressure is enough the display shows Running Now the Labaqua system is filling the tank with Grade II water As soon as the tank is full the system...

Page 14: ...le duration time 6 TOC level reduction and monitoring Total organic carbon TOC describes level of organic contaminants in pure water High concentration of organic contaminants may affect results of se...

Page 15: ...lation 7 Specifications The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 15 C to 30 C with maximum change at less than 2 C in 24 hour...

Page 16: ...e pump 50 L BS 070102 DK Water storage tank Comfort with dispense pump 100 L BS 070102 FK Water storage tank Comfort with dispense pump 200 L BS 070102 GK Water storage tank Comfort with dispense pump...

Page 17: ...key Fig 4 3 two times to select and confirm Switch off the system Wait 5 seconds and switch on the system again 8 5 Expendable parts and the replacement interval are shown below Description Replacemen...

Page 18: ...chnical support section on our website at the link below 9 3 Warranty covers only the units transported in the original package 9 4 If any manufacturing defects are discovered by the Customer an unsat...

Page 19: ...19 10 Declaration of Conformity...

Page 20: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 06 April 2017...

Reviews: