Biosan Labaqua Operating Manual Download Page 12

12 

4.5. 

Reconnect the outlet tube to the outlet John Guest elbow. 

4.6. 

Press 

Run

 key and wait until storage tank is filled with water.  

4.7. 

Put the end of 3/8’’ tank “Out” tube into a sink, press “Pump Grade 2” and dispense 
water from the tank. 

4.8. 

Let the system fill the tank again. As the tank is full, ask the customer to leave the 
system  overnight  (shut  off  tap  water  supply  for  safety  reasons).  The  system  will 
recirculate water in the tank and get rid of organic compounds as well as rinse the 
polishing module. 

4.9. 

The  microfilter  (or  ultrafilter)  should  be  rinsed  prior  to  use!  Ask  the  customer  to 
press “Pump Grade 1” next morning and dispense water in the tank through point-
of-use  filter  (microfilter  on  Labaqua  HPLC  or  ultrafilter  on  Labaqua  Bio).  The 
microfilter (or ultrafilter) needs to be rinsed with at least 20 litres of purified water. 

 

Fig. 3. Deionization and polishing module sites. The grey fittings above are used for 

connecting the modules. 

5.  Operation 

5.1. 

Switch  on  the  system  with  the  power  switch.  The  display  (Fig.  4/1)  lights  up  after 
few seconds. 

5.2. 

The display shows information in several rows: 

system messages and current status; 

error messages, if present; 

Status messages 

Description 

Standby 

Labaqua is switched off and does not produce water to the 
tank. 

LowPres 

Input pressure is too low. 

TnkFull 

The storage tank is full. 

Disp. 

Water is being dispensed. 

Recirculate 

Water in the storage tank is being recirculated 

 

 

 

Summary of Contents for Labaqua

Page 1: ......

Page 2: ...2...

Page 3: ...4 2 General information 5 3 Getting started 8 4 Rinsing the system 11 5 Operation 12 6 TOC level reduction and monitoring 14 7 Specifications 15 8 Cleaning and maintenance 17 9 Warranty and Claims 18...

Page 4: ...ign of the unit Do not block any ventilation openings Do not operate the unit with its covers removed Do not drink deionized water WATER LEAKAGE SAFETY Make sure that all water connection tubes are un...

Page 5: ...cific atmosphere Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriat...

Page 6: ...omatically switch on the boost pump Fig 1 4 and LowPres message will be shown on the display The boost pump Fig 1 4 is used to maintain pressure at the level necessary for the efficient operation of t...

Page 7: ...s and Grade I water is dispensed via the 0 22 m point of use microfilter Labaqua Bio system has an ultrafilter instead of the microfilter To dispense ISO 3696 Grade II water press the Pump Grade II ke...

Page 8: ...ank with pump level switch and tap 1 set 3 8 with elbow 3m overflow tube 1 pce 3 8 1m tank dispense tube 1 pce tank cable 1 pce hex key for controller module 1 pce Operation manual declaration of conf...

Page 9: ...the new pre filter under warranty only if we get a photo of the sediment filter installed in the tap water line Feed water hardness does not affect produced water quality However operation of the syst...

Page 10: ...move the packing material inside 3 4 8 Connect 1 4 John Guest outlet OUT to the inlet In of the tank 3 4 9 Connect 1 4 John Guest outlet Rec Out to the inlet Rec In of the tank 3 4 10 Connect 1 4 John...

Page 11: ...lled site push the tubes into John Guest fittings at the inner wall Then attach the module fully to the inner wall by putting the latch at the bottom of the module into slot on the wall 3 6 3 Screw th...

Page 12: ...next morning and dispense water in the tank through point of use filter microfilter on Labaqua HPLC or ultrafilter on Labaqua Bio The microfilter or ultrafilter needs to be rinsed with at least 20 li...

Page 13: ...is neither full nor disconnected and the feed water pressure is enough the display shows Running Now the Labaqua system is filling the tank with Grade II water As soon as the tank is full the system...

Page 14: ...le duration time 6 TOC level reduction and monitoring Total organic carbon TOC describes level of organic contaminants in pure water High concentration of organic contaminants may affect results of se...

Page 15: ...lation 7 Specifications The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 15 C to 30 C with maximum change at less than 2 C in 24 hour...

Page 16: ...e pump 50 L BS 070102 DK Water storage tank Comfort with dispense pump 100 L BS 070102 FK Water storage tank Comfort with dispense pump 200 L BS 070102 GK Water storage tank Comfort with dispense pump...

Page 17: ...key Fig 4 3 two times to select and confirm Switch off the system Wait 5 seconds and switch on the system again 8 5 Expendable parts and the replacement interval are shown below Description Replacemen...

Page 18: ...chnical support section on our website at the link below 9 3 Warranty covers only the units transported in the original package 9 4 If any manufacturing defects are discovered by the Customer an unsat...

Page 19: ...19 10 Declaration of Conformity...

Page 20: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 06 April 2017...

Reviews: