Biosan Labaqua Operating Manual Download Page 4

1.  Safety precautions 

The following symbol means: 

 

Caution!

 

Make sure you have fully read and understood the present  Manual 
before using the equipment. Please pay special attention to sections 
marked by this symbol. 

 

Caution!

 

Do  not  service  the  unit  while  the  open  UV  lamp  is  switched  ON. 
Otherwise  it  can  expose  the  operator  to  a  dangerous  level  of  UV 
emission. 

GENERAL SAFETY 

 

Use only as specified in the operating manual provided. 

 

The unit should not be used if dropped or damaged. 

 

The  unit  must  be  stored  and  transported  in  a  horizontal  position  (see  package 
label). 

 

After transportation or storage, keep the unit at room temperature for 2-3 hrs before 
connecting it to the electric circuit. 

 

Use  only  cleaning  and  decontamination  methods  recommended  by  the 
manufacturer. 

 

Use only original accessories (fuses, tubes, etc.) provided by the manufacturer and 
ordered specifically for this model. 

 

Do not make modifications to the design of the unit. 

 

Do not block any ventilation openings. 

 

Do not operate the unit with its covers removed. 

 

Do not drink deionized water. 

WATER LEAKAGE SAFETY 

 

Make sure that all water connection tubes are unbent, kink-free. 

 

Make sure that all water connectors are tight. 

 

When installing the tank, connect the fitting marked „Overflow” to the drain. Make 
sure  that  the 

drain level is lower than the „Overflow” fitting level. This will prevent 

water leakage in case of tank level sensor failure. 

 

Do not operate the tank pump if there is no water in the tank. Tank pump may fail if 
it is run dry. 

 

A  5 

μm sediment filter has to be installed in the feed water supply line. Failure to 

install  the  filter  may  result in clogging  of  strainer  inside  the  system  and  water  flow 
blockage. 

 

Shut off tap water supply when leaving the system unattended. 

ELECTRICAL SAFETY 

 

A  power  line  regulator  may  be  required  in  the  areas  where  the  supplied  power  is 
subject to voltage fluctuations exceeding 10% of the nominal. 

 

Connect  only  to  a  mains  outlet  with  voltage  corresponding  to  that  on  the  serial 
number label. 

 

Verify that the fuses are properly installed. 

Summary of Contents for Labaqua

Page 1: ......

Page 2: ...2...

Page 3: ...4 2 General information 5 3 Getting started 8 4 Rinsing the system 11 5 Operation 12 6 TOC level reduction and monitoring 14 7 Specifications 15 8 Cleaning and maintenance 17 9 Warranty and Claims 18...

Page 4: ...ign of the unit Do not block any ventilation openings Do not operate the unit with its covers removed Do not drink deionized water WATER LEAKAGE SAFETY Make sure that all water connection tubes are un...

Page 5: ...cific atmosphere Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms BIOLOGICAL SAFETY It is the user s responsibility to carry out appropriat...

Page 6: ...omatically switch on the boost pump Fig 1 4 and LowPres message will be shown on the display The boost pump Fig 1 4 is used to maintain pressure at the level necessary for the efficient operation of t...

Page 7: ...s and Grade I water is dispensed via the 0 22 m point of use microfilter Labaqua Bio system has an ultrafilter instead of the microfilter To dispense ISO 3696 Grade II water press the Pump Grade II ke...

Page 8: ...ank with pump level switch and tap 1 set 3 8 with elbow 3m overflow tube 1 pce 3 8 1m tank dispense tube 1 pce tank cable 1 pce hex key for controller module 1 pce Operation manual declaration of conf...

Page 9: ...the new pre filter under warranty only if we get a photo of the sediment filter installed in the tap water line Feed water hardness does not affect produced water quality However operation of the syst...

Page 10: ...move the packing material inside 3 4 8 Connect 1 4 John Guest outlet OUT to the inlet In of the tank 3 4 9 Connect 1 4 John Guest outlet Rec Out to the inlet Rec In of the tank 3 4 10 Connect 1 4 John...

Page 11: ...lled site push the tubes into John Guest fittings at the inner wall Then attach the module fully to the inner wall by putting the latch at the bottom of the module into slot on the wall 3 6 3 Screw th...

Page 12: ...next morning and dispense water in the tank through point of use filter microfilter on Labaqua HPLC or ultrafilter on Labaqua Bio The microfilter or ultrafilter needs to be rinsed with at least 20 li...

Page 13: ...is neither full nor disconnected and the feed water pressure is enough the display shows Running Now the Labaqua system is filling the tank with Grade II water As soon as the tank is full the system...

Page 14: ...le duration time 6 TOC level reduction and monitoring Total organic carbon TOC describes level of organic contaminants in pure water High concentration of organic contaminants may affect results of se...

Page 15: ...lation 7 Specifications The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 15 C to 30 C with maximum change at less than 2 C in 24 hour...

Page 16: ...e pump 50 L BS 070102 DK Water storage tank Comfort with dispense pump 100 L BS 070102 FK Water storage tank Comfort with dispense pump 200 L BS 070102 GK Water storage tank Comfort with dispense pump...

Page 17: ...key Fig 4 3 two times to select and confirm Switch off the system Wait 5 seconds and switch on the system again 8 5 Expendable parts and the replacement interval are shown below Description Replacemen...

Page 18: ...chnical support section on our website at the link below 9 3 Warranty covers only the units transported in the original package 9 4 If any manufacturing defects are discovered by the Customer an unsat...

Page 19: ...19 10 Declaration of Conformity...

Page 20: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 06 April 2017...

Reviews: