Appendix A.
M
anufacturer’s declaration - electromagnetic
immunity
The
FC1400
system is intended for use in the electromagnetic environment specified below.
The customer or the user of the
FC1400
system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood, conc
rete or ceramic tile. If floors
are covered with synthetic m
aterial, the relative humidity
should be at least 30 %
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply li
nes 1kV for input/outp
ut lines
2kV for power suppl
y lines
1kV for input/output l
ines
Mains power quality should
be that of a typical commerci
al or hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mod
e
2 kV common mode
Mains power quality should
be that of a typical commerci
al or hospital environment.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic fi
elds should be at levels char
acteristic of a typical location
in a typical commercial or ho
spital environment.
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
т (>95% dip in
U
т) for 0.5cycle
40%
U
т (60% dip in
U
т
) for 5 cycle
70%
U
т (30% dip in
U
т
) for 25 cycle
<5%
U
т (<95% dip in
U
т ) for 5 s
<5%
U
т (>95% dip i
n
U
т) for 0.5cycle
40%
U
т (60% dip in
U
т ) for 5 cycle
70%
U
т (30% dip in
U
т) for 25 cycle
<5%
U
т (<95% dip i
n
U
т ) for 5 s
Mains power quality should
be that of a typical commerci
al or hospital environment. If
the user of the system requir
es continued operation durin
g power mains interruptions,
it is recommended that the
system be powered from an
uninterruptible power supply
or a battery
Note:
U
т is the a.c. mains voltage prior to application of the test level.