FC1400 User Manual
FC1400OM-2.04 59 / 95
FC1400 Operation Manual
3) Visual Alarm
Alarm messages appear in the Alarm Status Area on the screen. If more than one alarm occurred,
messages are changed every two seconds, with an arrow displayed next to the messages. The colors
of the alarm messages match the levels of the alarms. Medium Alarms are displayed in red, Low
Alarms, in yellow, and product condition alarms, in sky blue. The asterisks next to alarm messages
also match the levels of alarms. ** indicates medium alarms, * indicates Low Alarms, and no asterisk
indicates equipment condition alarms.
When a patient condition alarm occurs, the color of the numerical value in the numeral window will be
changed to red.
Visual Alarms will be maintained even after identifying the alarm statuses.
4) Alarm LED
There are two types of Alarm LED around the exterior to help checking the alarm condition from far.
Yellow LED: Alarm occurred due to patient condition.
Medium Alarm: Flickering once a second.
Low Alarm: Stay on
Green LED: Alarm occurred due to Technical Alarm and stays on.
5) Audible Silence and Pause
When various alarms occur during the operation, this function is used to silence or pause them, while
the user is verifying the message. If
“Alarm silence” button is clicked one time, the device will go
silence for 1 minute. If a new alarm occurs during the period, alarm sound will ring again. If
“Alarm
silence
” button is clicked twice, the device will go into “Pause” state for 5 minutes. Under “Pause” no
alarm will sound even if a new alarm is generated. If
“Alarm Silence” button is clicked three times, any
sound alarm will be turned off and no sound will be made even if a new alarm occurs.
Visual alarm will be displayed until the situation is resolved.
To silence the alarm, touch
“Alarm silence” button on control panel.
: Alarm Silence Button
Warning
When monitoring patient conditions, do not rely solely on audible signals. Reducing or
turning off the volume of audible signals may endanger the patient.