FC1400 User Manual
FC1400OM-2.04 80 / 95
FC1400 Operation Manual
Chapter 19. Product specification
General specification
Dimension
296(w)x.305.5(H)x97.5(D)(Approx. 2.9Kg)
Display
7” Wide (800 X 480)
Recorder
Method : Thermal Array Print
Type : Roll type
Print speed : 1,2,3cm/min,( Real time)
30 cm/min (Trace, 2,4 cm/min setting)
20 cm/min (Trace, 1 cm/min setting)
Paper feeding function
Performance Specification
Fetal Heart Rate
Input Signal : Ultrasound pulsed doppler
FHR detection method : Autocorrelation
FHR range : 50~210
FHR accuracy : 120~160 : ±1 bpm
Except 120~160 : ±2 bpm
Ultrasound Transducer
Operating mode : PWD Mode
Transducer Type : 7-crystal
Ultrasound frequency : 1.0MHz
Pulse Repetition Frequency : 3125Hz
Spatial-Peak Temporal Average Intensity : <10mW/cm2
Uterine Contraction
Input Source : External Transducer
Reference Control : One touch switch
Auto zeroing
Measurement range : 0~99
Auto CTG Analysis
Average Baseline FHR
Number of TOCO
Number of Acceleration
Number of Deceleration : Late, Early, Variable
High/Low Episode
Short term Variability
Signal Loss
* CTG Analysis results is printed out every 10 minutes
Data Storage
Storage for 72 Hours
Power Specification
Power
DC Input: 18V , 2.8A / Adaptor “USE THE ONLY BridgePower Corp
BPM050S18F02”
Battery(Option)
Li-ion : 4 hours ( charging)
2 hours(discharging)
External Link
LAN, Wi-fi, USB,SD
Default Alarm Setting
Alarm Parameter
US1/ US2
ON
Alarm Limit
US1/US2
160
120
Alarm Level
US1/US2
Medium