FC1400 User Manual
FC1400OM-2.04 11 / 95
FC1400 Operation Manual
CAUTIONS
Instruction for Use
For continued safe use of this equipment, it is necessary that the instructions are followed. However,
instructions listed in this manual in no way supersede established medical practices concerning
patient care.
Loss of Data
Should the monitor at any time temporarily lose patient data, the potential exists that active monitoring
is not being done. Close patient observation or alternate monitoring devices should be used until
monitor function is restored.
If the monitor does not automatically resume operation within 60 seconds, power cycle the monitor
using the power on/off switch. Once monitoring is restored, you should verify correct monitoring state
and alarm function.
Maintenance
Regular preventive maintenance should be carried out annually (Technical inspections). You are
responsible for any requirements specific to your country.
MPSO
The use of a multiple portable socket outlet (MPSO) for a system will result in an enclosure leakage
current equal to the sum of all individual earth leakage currents of the system if there is an interruption
of the MPSO protective ground conductor. Do not use an additional extension cable with the MPSO as
it will increase the chance of the single protective ground conductor interruption.
Negligence
BIONET does not assume responsibility for damage to the equipment caused by improperly vented
cabinets, improper or faulty power, or insufficient wall strength to support equipment mounted on such
walls.