Biomedical Life Systems BioStim NMS2 Instructions Manual Download Page 6

 

2.

 

Introduction BioStim® NMS2 

 

2.1 Precautions should be taken when stimulation is used;  

•  After recent surgical procedure where muscle contraction may 

disrupt the healing process. 

• After an acute trauma or fracture where there is a tendency to 

hemorrhage. 

• Over the menstruating uterus. 

• Where the sensory nerve damage has caused the loss of normal 

skin sensation.  

EQUIPMENT not suitable for use in the presence of a FLAMMABLE 

ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS 

OXIDE. 

 

2.2 Caution: 

NMS or EMS Devices should only be used under medical supervision 

for adjunctive therapy in the treatment of medical diseases and 

conditions. Dependent upon government regulation, this device 

may or may not require a medical prescription. Federal law (USA) 

restricts the sale by, or on the order of, a physician so licensed by 

the State. Keep out of reach of children. 

 

2.3 Adverse Reactions: 

Some patients may experience skin irritation or hypersensitivity due 

to electrical stimulation or electrically  conductive medium. The 

irritation can usually be reduced by using an alternative conductive 

medium, or alternate electrode placement. Improper use of 

stimulation may result in skin irritation and burns beneath the 

electrodes. Electrode placement and stimulation settings should be 

based on the guidance of the prescribing practitioner. 

 

2.5 Contraindications:

  

 

1) Powered muscle stimulators should not be used on patients with 

cardiac demand pacemakers. 

2) NMS or EMS devices should not be applied over, or in proximity 

to, cancerous lesions. 

Summary of Contents for BioStim NMS2

Page 1: ..._________ _______________________ _______________________ _______________________ _______________________ _______________________ _______________________ _______________________ ______________________...

Page 2: ...Set Therapy Mode Parameters 5 Batteries 11 6 Electrodes and Lead Wires 12 6 1 Connect Electrodes and Leadwires 7 Graphic Symbol Definitions 13 7 1 BioStim NMS Button Descriptions 8 Programming Instru...

Page 3: ...Setting the Reciprocal RECIPR Mode 10 Using Therapy Modes 1 8 27 11 Malfunctions 28 12 Safety and Technical Checks 29 13 Maintenance and Care 29 14 Warranty 29 BIOSTIM NMS2 INSTRUCTION MANUAL Please c...

Page 4: ...3...

Page 5: ...ention or Retardation of Muscle Disuse Atrophy Muscle disuse atrophy is a reduction in muscle tone and size due to prolonged impairment or joint immobility from surgery or injury The use of NMS to con...

Page 6: ...regulation this device may or may not require a medical prescription Federal law USA restricts the sale by or on the order of a physician so licensed by the State Keep out of reach of children 2 3 Adv...

Page 7: ...as Stimulation should not be applied over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Persistent use of stimulation in the presence of skin i...

Page 8: ...biphasic square SYM and Asymmetrical biphasic square ASYM Pulse Rate Adjustable 1 120 Hz Hertz pulses sec Pulse Width Adjustable 50 400 S microseconds Ramp On 0 99 seconds Ramp Off 0 99 seconds On Tim...

Page 9: ...ts internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emission CISPR 11 Class B This device is suitable for use in...

Page 10: ...Mode or a Pulse Rate greater than 60Hz with a symmetrical biphasic square waveform should be used for the relaxation of muscle spasms only Use of constant stimulation and or rates above the normal phy...

Page 11: ...tions In Seconds Therapy 1 Cycled Ramp On Off 2 2 Symmetrical Biphasic Wave On Time 10 PR 35Hz PW 300 S Off Time 20 Therapy 2 Cycled Ramp On Off 2 2 Asymmetrical Biphasic Wave On Time 10 PR 35Hz PW 40...

Page 12: ...ottom right hand corner of the LCD display screen When this image appears the batteries should be changed to ensure maximum performance If you have saved programs into the device they will remain stor...

Page 13: ...ectrodes and leadwires should be reordered through the dealer who supplied the device 6 1 Connect electrodes and lead wires 1 Make sure the device is turned off if the device is on press the button la...

Page 14: ...MPLITUDE CONTROL 4 9 INCREASE VALUE DECREASE VALUE 5 MODE SELECT 6 ENTER 7 SETTING INFORMATION KEY 8 PULSE RATE PULSE WIDTH WAVEFORM SELECT 10 TIMER KEY Refer to operating instructions An IEC 601 1 sa...

Page 15: ...the device Button 6 Enter This key is used to accept a value Button 7 Setting Information Key This key is used to view the timing options RAMP TIME ON TIME OFF TIME DELAY of a Mode or Therapy without...

Page 16: ...the 2 or 3 keys The 2 key controls the amplitude of Channel 1 and the 3 key controls the amplitude of Channel 2 In CONST CYCLED and RECIPR modes you may adjust the settings before beginning stimulati...

Page 17: ...ME value lesser than the sum of the RAMP ON and RAMP OFF values as the ON TIME is the total time the device intensity is on in the cycle including the ramp times Note The ON TIME value is set independ...

Page 18: ...l 1 and Channel 2 total cycle times must be the same The following is an example of a Cycled Program with a DELAY time of 5 seconds Channel 1 Channel 2 Ramp on 2 0 sec 2 0 sec Ramp off 2 0 sec 2 0 sec...

Page 19: ...ing the Enter Key 6 until there are no blinking values Step 7 Optional To review the timing option settings without changing them use the Info Key 7 to scroll through the settings Step 8 You may begin...

Page 20: ...blinking on the screen press the Enter Key 6 until there are no blinking values 8 8 E 6 4 9 8 5 Patient Timer The BioStim NMS2 device has a patient Timer feature When the timer is enabled the device c...

Page 21: ...9 10 E 6 8 6 Patient Lock The BioStim NMS2 has a Patient Parameter Lock Feature When activated this feature locks all of the displayed settings except the Intensity Controls mA ON OFF capabilities and...

Page 22: ...ompliance Meter read out See Section 8 7 Patient Compliance Meter The Patient Lock Feature is now deactivated 8 7 Patient Compliance Meter The Patient Compliance Meter records usage time in hours begi...

Page 23: ...n reached a bar next to the 1 on the left Amplitude Display F will appear Each bar indicates 100 hours of use The maximum amount of time that will be displayed is 1099 hours The device does not reset...

Page 24: ...ttach electrodes to the body per the instructions given by the prescribing practitioner Step 4 Turn device on by pressing the ON OFF Key 1 Step 5 Press the Mode Key 5 until CONST appears on the screen...

Page 25: ...same When ON TIME OFF TIME values are different between Channel 1 and 2 or when a DELAY value is set the device automatically calculates and adjusts the OFF TIME of Channel 2 to ensure the total Cycl...

Page 26: ...3 Changing the Timing Options RAMP TIME ON TIME OFF TIME DELAY Step 7 Optional To change the Hz S or waveform See section 8 4 Changing PR Hz PW US and Waveform Values Step 8 Optional To set the Patie...

Page 27: ...he instructions on the electrode packaging Step 3 Attach electrodes to the body per the instructions given by the prescribing practitioner Step 4 Turn device on by pressing the ON OFF Key 1 Step 5 Pre...

Page 28: ...at any time by pressing the 2 or 3 buttons in the direction 10 Using THERAPY Modes 1 8 There are eight pre programmed therapies to choose from in the BioStim NMS2 device These therapies are fixed and...

Page 29: ...tions occur while using this device check Whether the leadwires and electrodes are correctly connected to the device The leadwires should be inserted firmly into the device sockets For possible damage...

Page 30: ...ntact your dealer 13 Maintenance and Care The case housing is made of ABS plastic and can be cleaned with isopropyl alcohol NOTE Do not smoke or work with an open flame for example candles etc when wo...

Page 31: ...er purchaser BioMedical Life Systems Inc shall not be liable for incidental or consequential damages resulting from the sale or use of the device In the USA some states do not allow the exclusion or l...

Reviews: