![Biomedical Life Systems BioStim NMS2 Instructions Manual Download Page 18](http://html1.mh-extra.com/html/biomedical-life-systems/biostim-nms2/biostim-nms2_instructions-manual_2753985018.webp)
17
DELAY:
The amount of time in seconds between the begininning
of Channel 1 cycle and beginning 2 cycle. (DELAY is available in
the CYCLED mode only). The DELAY can be adjusted from 1-99
seconds.
Note:
If a "DELAY" value is set in the CYCLED mode, the number
of seconds set for the DELAY will be automatically deducted from
the OFF TIME of Channel 2, as both Channel 1 and Channel 2 total
cycle times must be the same. The following is an example of a
Cycled Program with a DELAY time of 5 seconds.
Channel 1
Channel 2
Ramp on
2.0 sec.
2.0 sec.
Ramp off
2.0 sec.
2.0 sec.
On time
20 sec.
10 sec.
Off time
10 sec.
15 sec.
Delay
N/A
5 sec.
TOTAL
35 sec.
35 sec.
8.3: Changing Timing Options; RAMP TIME, ON TIME OFF