RAMP ON:
Is the amount of time in seconds the device itensity
ramps up from the OFF Time to the maximum intensity of the ON
Time in each cycle.
RAMP OFF:
Is the amount of time in seconds the device intensity
ramps down from the ON Time to the OFF Time in each cycle.
Special Notes Regarding RAMP ON/OFF Times
•
In the BioStim® NMS
2
the RAMP ON can be different from the
RAMP OFF value in the cycle.
•
The RAMP ON and RAMP OFF values set to both Channel 1
and Channel 2. ON TIME: The total amount of time seconds in
each cycle that the device intensity is on, including the ramp
times.
Special Notes Regarding ON TIME:
•
Once the RAMP ON and RAMP OFF values have been set, the
device will not allow the user to set a ON TIME value lesser than
the sum of the RAMP ON and RAMP OFF values, as the ON
TIME is the total time the device intensity is on in the cycle,
including the ramp times.
•
Note: The ON TIME value is set independently for Channel 1
and Channel 2 and may be a different value for Channel 1 than
Channel 2.
•
If different ON TIME values are set for Channel 1 and Channel 2
in a CYCLED program, the difference between the Channel 1
and Channel 2 ON TIME will be automatically added or
subtracted from the Channel 2 OFF TIME as the total cycle
times for Channel 1 and Channel 2 need to be the same in the
CYCLED mode. In the RECIPR mode, the total cycle time does
not need to be the same, therefore, the user may set different
ON TIME values for Channel 1 and Channel 2 without the
device deducting the difference from the OFF TIME of Channel
2.
OFF TIME:
The amount of time in seconds that the device
intensity is off in each cycle.