![Biomedical Life Systems BioStim NMS2 Instructions Manual Download Page 28](http://html1.mh-extra.com/html/biomedical-life-systems/biostim-nms2/biostim-nms2_instructions-manual_2753985028.webp)
27
Step 6:
(Optional) To Change the Timing options refer to sec
tion(s) 8.2-8.3 (Changing the Timing Options; RAMP TIME, ON
TIME OFF TIME & DELAY)
Step 7:
(Optional) To change the Hz, μS or waveform; See
section 8.4 (Changing PR (Hz), PW (US) and Waveform Values)
Step 8:
(Optional) To set the Patient Timer see section 8.5
(Patient Timer). If the timer is not set, the device will continue to
run until the "ON/OFF" key (1)
is pressed, or if the device is
idol (with no intensity settings) for 7 consecutive minutes.
Step 9:
To begin stimulation; increase the intensity of each
Channel by pressing the (2)
or (3)
buttons in the "+"
direction. You may also decrease the intensity at any time by
pressing the (2)
or (3)
buttons in the "-" direction.
10.
Using THERAPY Modes 1-8
There are eight pre-programmed therapies to choose from in the
BioStim NMS2 device. These therapies are "fixed" and cannot be
altered. To view the settings refer to section 4.2 (Pre-Set Therapy
Mode Parameters).
The Pre-Set Therapies are designed for ease of use. The most
common CONSTANT, CYCLED, and RECIPROCAL settings are
preprogrammed into the device and labeled THERAPY 1-8. If
adjustments need to be made to the settings, the Programma ble
Modes will need to be used instead of a pre-set Therapy. (See
Sections 9-9.6 Programmable Modes Explained).
STEP 1:
Turn device on by pressing the "ON/OFF" Key (1)
STEP 2:
Press the "Mode" Key (5)
repeatedly to scroll
through the different Mode Selections; CONST., CYCLED, RECIPR.,
THERAPY1, THERAPY 2, THERAPY 3, THERAPY 4, THERAPY
Summary of Contents for BioStim NMS2