4.2 Pre-Set Therapy Mode Parameters:
There are 8 Pre-Set Therapies in the BioStim® NMS².
Note: Settings in the following THERAPY modes are FIXED and
cannot be changed. If custom settings are needed, please see
Programmable Modes Explained; (Section 9)
Therapy
Stimulation Pattern;
(Const, Cycled or Reciprocal)
Timing Options
(In Seconds)
Therapy
1
Cycled
Ramp On/Off
2/2
Symmetrical Biphasic Wave
On Time
10
PR 35Hz PW
300μS
Off Time
20
Therapy
2
Cycled
Ramp On/Off 2/2
Asymmetrical Biphasic Wave
On Time
10
PR
35Hz PW 4
00μS
Off Time
20
Therapy
3
Cycled
Ramp On/Off 2/2
Symmetrical Biphasic Wave
On Time
10
PR
50Hz PW 180
μS
Off Time
20
Therapy
4
Cycled
Ramp On/Off 2/2
Symmetrical Biphasic Wave
On Time
5
PR
20Hz PW
250
μS
Off Time
10
Therapy
5
Constant
Ramp On/Off
0
Symmetrical Biphasic Wave
On Time
0
PR
10Hz
PW
5
0μS
Off Time
0