10.1. Cleaning the tip cone

1.  Remove the ejection collar: Gently twist it anti-

clockwise and slide off (Fig. 9). Remove 

fi

 lter i 

fi

 tted.

2.  Using Biohit Proline Biocontrol or ethanol and 

soft lintfree cloth, disinfect the tip ejector collar 

and the tip cone.

3.  Unscrew, anti-clockwise, the nose cone and 

remove it, exposing the piston (Fig.10). If the 

piston is sticked to the tip cone remove it with a 

pair of tweezers.

4.  To avoid scratching the surface of the piston 

use Biohit Proline Biocontrol or ethanol and a 

lint-free tissue when cleaning the piston. Let the 

parts dry.

5.  Grease the piston thinly with the grease provided. 

Do not use any other grease. Check that no lint or 

particles are on the surface of the piston. Avoid 

excess  grease,  especially  at  the  bottom  of  the 

piston.

Note:

 For complete decontamination place the tip 

cone, tip ejector collar, piston, O-ring and spring into a beaker containing 

Biohit Proline Biocontrol and leave for at least 30 minutes, rinse the parts 

with distilled water, then dry preferably with warm air. 

 

6.  Reassemble the pipette by screwing the piston in its place and replace 

the tip ejector collar allowing the ejector handle connection to snap into 

the attachment notch of the ejector collar. Replace new 

fi

 lter i 

fi

 tted.

7.  Test the pipette by pressing the START-button severel times. Test the 

tip ejector operation.

Note:

 Check the performance of your Biohit Proline pipette regularly e.g. 

every 3 months and always after in-house service or maintenance.

10.2. Battery Replacement

If the battery does not hold a suf

fi

 cient charge for proper operation of the 

pipette, follow these steps for replacement of the battery.

1.   Switch the unit “OFF”.

2.   Remove the top two screws on the back of the pipette and remove the 

battery cover (Fig. 11).

3.   Carefully remove the battery by lifting it straight out of the holder.

Fig. 9

Fig. 10

Summary of Contents for Proline 710010

Page 1: ...Proline K ytt ohje Bruksanvisning Instruction Manual...

Page 2: ...K ytt ohje 1 Bruksanvisning 23 Instruction Manual 45 Specifications 67 Proline...

Page 3: ......

Page 4: ......

Page 5: ...nd Ejecting tips 52 5 6 Optional Filters 52 6 PROGRAMMING THE PIPETTOR 53 6 1 Mode Selection and Mode Recall 53 6 2 Setting Speeds 53 6 3 Pipetting Mode P 54 6 4 Reverse Pipetting Mode rP 54 6 5 Multi...

Page 6: ...piston movements Also selected models offer replaceable tip cone filters to help prevent contamination and damage Its light weight and ergonomic controls take the effort out of pipetting to help redu...

Page 7: ...100 l 1 l 300 l 350 l 710320 12 kan 25 250 l 5 l 350 l 710810 12 kan 50 1200 l 10 l 1200 l 2 3 Biohit tips The full range of Biohit tips are recommended forusewithBiohitpipettes Thesedetachable dispos...

Page 8: ...nce card Performance certificate in accordance with ISO 8655 6 Make sure that all items are included and that no damage has occured during shipment Note The pipette can only be charged with the Biohit...

Page 9: ...keep the pipette in the stand for 12 hours to fully charge the pipette before continuing use 4 Display shows Press the START button twice and the pipette is ready for the default pipetting at maximum...

Page 10: ...r level POM Display glass Finger support back ABS Keyboard silicon Start button POM Keyboard silicon Start button POM Handle ASA PC Handle ASA PC Tip cone housing ASA PC Tip ejector bar PETP Tip cone...

Page 11: ...ton If the START button is kept down the piston will stop in the lowest position until the button is released This feature is applicable to all modes of operation except for multiple dispensing d 5 3...

Page 12: ...the use of a removable filter as an option The filter prevents liquids and liquid vapours from entering the pipette The filter does not affect the calibration of the pipette Biohit filter forceps shou...

Page 13: ...n only be activated when the piston is in its home position arrow right sign is lit not in the middle of aspiration or dispensing actions 6 2 Setting Speeds all models 1 Press to display the current a...

Page 14: ...y the excess volume remains in the tip and is discarded 1 Press until is displayed 2 Press to confirm mode change 3 Use and to select the desired pipetting volume 4 Press to confirm selection 5 Positi...

Page 15: ...d press the START button The sign and ARROW LEFT are lit to indicate the reset function 8 Position the tip to discard priming excess and press the START button 9 To dispense position the tip press the...

Page 16: ...irm selection 7 Position the tip to aspirate volume 1 and then press the START button 8 With the tip in the air press the START button again to aspirate an air gap 9 Position the tip to aspirate volum...

Page 17: ...ing 5 Position the tip to aspirate volume 1 and then press the START button 6 With the tip in the air press the START button to aspirate an air gap 7 Position the tip to aspirate volume 2 and press th...

Page 18: ...p to aspirate and press the START button 8 Position the tip to discard priming excess and press the START button again 9 Position the tip to dispense and press the START button Repeat the action until...

Page 19: ...s are made possible Two of the modes P and dd feature an automatic blow out and others leave an excess liquid in the tip The user should follow the recommendations below to ensure optimal performance...

Page 20: ...rinse the tip and change the tip after each pipetting Check the pipette tip and liquid are at room temperature Avoid contaminating the tip cone do not rest the pipette on its side Always place the pip...

Page 21: ...display 2 Confirm the mode change by pressing 3 Select the desired pipetting volume to be calibrated by using to increase and to decrease 4 Press to confirm the selection 5 Press and simultaneously to...

Page 22: ...ewet the new tip once 4 Aspirate the test water immersing the tip only 2 3 mm below the surface of the water and keeping the pipette vertical Withdraw the pipette vertically and touch the tip against...

Page 23: ...p cone tip ejector collar piston O ring and spring into a beaker containing Biohit Proline Biocontrol and leave for at least 30 minutes rinse the parts with distilled water then dry preferably with wa...

Page 24: ...the error message please do the following Note As this procedure will empty the tip it is recommended that you remove the tip before resetting the pipette 1 Place the pipette in the charging stand fo...

Page 25: ...piston and O ring Instrument damaged Attach new tip Attach firmly Use original tips Clean the tip cone attach new tip Clean and grease piston and tip cone Grease accordingly Send for service Pipette b...

Page 26: ...oline Electronic pipette is tested before shipping by the manufacturer The Quality Assurance Procedure is your guarantee that the Biohit Proline Electronic pipette you have purchased is ready for use...

Page 27: ...10 00 2 50 710010 1 ch 5 100 l 100 l 50 l 10 l 5 l 0 40 0 70 2 00 2 50 0 15 0 30 1 00 1 80 2 20 710030 1 ch 10 250 l 250 l 125 l 25 l 10 l 0 40 0 60 2 00 2 00 0 15 0 20 1 00 1 00 2 25 710100 1 ch 10 5...

Page 28: ...10 l 5 l 0 50 0 70 2 50 4 00 0 20 0 30 1 50 2 50 2 20 710220 8 ch 25 250 l 250 l 125 l 25 l 0 40 0 60 1 50 0 15 0 20 1 00 2 10 710800 8 ch 50 1200 l 1200 l 600 l 120 l 50 l 0 50 1 00 3 00 8 00 0 15 0...

Page 29: ...atenttihakemuksilla Biohit Proline on Biohit Oyj n rekister ity tavaramerkki P grund av kontinuerling utveckling f rbeh ller sig Biohit r tten till f r ndringar av specifikationer utan separat anm lan...

Reviews: