56

6.6. Diluting Mode (dd)

(all models, except 710800 and 710810)

Two different solutions separated by an air gap are aspirated and then 

dispensed together with automatic blow-out. The purpose of the air gap is 

to avoid contamination when aspirating the second volume but it will not 

prevent the two liquids from mixing in the tip.

1. Press 

 to display 

.

2. Con

fi

 rm the mode change by pressing 

.  

3.  Select the desired diluent volume (volume 1) by using 

 and 

.                         

4. Con

fi

 rm by pressing 

.  

5. Press 

 and 

 to select the sample volume (volume 2) on   the 

display.  

6. Press 

 to con

fi

 rm selection. 

7.  Position the tip to aspirate volume 1 and then press the START button.  

8.  With the tip in the air press the START button again to aspirate an air       

gap.  

9.  Position the tip to aspirate volume 2 and press the START button.  

10.  Finally, position the tip to dispense and press the START button.

6.7. Mixing Mode with Pipetting or Diluting (*)

(all models)

The piston is automatically moved up and down to mix the liquid in the 

delivery vessel. The mixing time is controlled by the START button.

1. Use 

 to select either 

 or 

 onto the display.  

2. Press 

 to switch on mixing. The display should read either 

 or 

. Please note that selecting 

 will alternately switch mixing 

on and off. 

3. Con

fi

 rm the mode change by pressing the 

 button.

Step 7.

Step 10.

Step 8.

Step 9.

Summary of Contents for Proline 710010

Page 1: ...Proline K ytt ohje Bruksanvisning Instruction Manual...

Page 2: ...K ytt ohje 1 Bruksanvisning 23 Instruction Manual 45 Specifications 67 Proline...

Page 3: ......

Page 4: ......

Page 5: ...nd Ejecting tips 52 5 6 Optional Filters 52 6 PROGRAMMING THE PIPETTOR 53 6 1 Mode Selection and Mode Recall 53 6 2 Setting Speeds 53 6 3 Pipetting Mode P 54 6 4 Reverse Pipetting Mode rP 54 6 5 Multi...

Page 6: ...piston movements Also selected models offer replaceable tip cone filters to help prevent contamination and damage Its light weight and ergonomic controls take the effort out of pipetting to help redu...

Page 7: ...100 l 1 l 300 l 350 l 710320 12 kan 25 250 l 5 l 350 l 710810 12 kan 50 1200 l 10 l 1200 l 2 3 Biohit tips The full range of Biohit tips are recommended forusewithBiohitpipettes Thesedetachable dispos...

Page 8: ...nce card Performance certificate in accordance with ISO 8655 6 Make sure that all items are included and that no damage has occured during shipment Note The pipette can only be charged with the Biohit...

Page 9: ...keep the pipette in the stand for 12 hours to fully charge the pipette before continuing use 4 Display shows Press the START button twice and the pipette is ready for the default pipetting at maximum...

Page 10: ...r level POM Display glass Finger support back ABS Keyboard silicon Start button POM Keyboard silicon Start button POM Handle ASA PC Handle ASA PC Tip cone housing ASA PC Tip ejector bar PETP Tip cone...

Page 11: ...ton If the START button is kept down the piston will stop in the lowest position until the button is released This feature is applicable to all modes of operation except for multiple dispensing d 5 3...

Page 12: ...the use of a removable filter as an option The filter prevents liquids and liquid vapours from entering the pipette The filter does not affect the calibration of the pipette Biohit filter forceps shou...

Page 13: ...n only be activated when the piston is in its home position arrow right sign is lit not in the middle of aspiration or dispensing actions 6 2 Setting Speeds all models 1 Press to display the current a...

Page 14: ...y the excess volume remains in the tip and is discarded 1 Press until is displayed 2 Press to confirm mode change 3 Use and to select the desired pipetting volume 4 Press to confirm selection 5 Positi...

Page 15: ...d press the START button The sign and ARROW LEFT are lit to indicate the reset function 8 Position the tip to discard priming excess and press the START button 9 To dispense position the tip press the...

Page 16: ...irm selection 7 Position the tip to aspirate volume 1 and then press the START button 8 With the tip in the air press the START button again to aspirate an air gap 9 Position the tip to aspirate volum...

Page 17: ...ing 5 Position the tip to aspirate volume 1 and then press the START button 6 With the tip in the air press the START button to aspirate an air gap 7 Position the tip to aspirate volume 2 and press th...

Page 18: ...p to aspirate and press the START button 8 Position the tip to discard priming excess and press the START button again 9 Position the tip to dispense and press the START button Repeat the action until...

Page 19: ...s are made possible Two of the modes P and dd feature an automatic blow out and others leave an excess liquid in the tip The user should follow the recommendations below to ensure optimal performance...

Page 20: ...rinse the tip and change the tip after each pipetting Check the pipette tip and liquid are at room temperature Avoid contaminating the tip cone do not rest the pipette on its side Always place the pip...

Page 21: ...display 2 Confirm the mode change by pressing 3 Select the desired pipetting volume to be calibrated by using to increase and to decrease 4 Press to confirm the selection 5 Press and simultaneously to...

Page 22: ...ewet the new tip once 4 Aspirate the test water immersing the tip only 2 3 mm below the surface of the water and keeping the pipette vertical Withdraw the pipette vertically and touch the tip against...

Page 23: ...p cone tip ejector collar piston O ring and spring into a beaker containing Biohit Proline Biocontrol and leave for at least 30 minutes rinse the parts with distilled water then dry preferably with wa...

Page 24: ...the error message please do the following Note As this procedure will empty the tip it is recommended that you remove the tip before resetting the pipette 1 Place the pipette in the charging stand fo...

Page 25: ...piston and O ring Instrument damaged Attach new tip Attach firmly Use original tips Clean the tip cone attach new tip Clean and grease piston and tip cone Grease accordingly Send for service Pipette b...

Page 26: ...oline Electronic pipette is tested before shipping by the manufacturer The Quality Assurance Procedure is your guarantee that the Biohit Proline Electronic pipette you have purchased is ready for use...

Page 27: ...10 00 2 50 710010 1 ch 5 100 l 100 l 50 l 10 l 5 l 0 40 0 70 2 00 2 50 0 15 0 30 1 00 1 80 2 20 710030 1 ch 10 250 l 250 l 125 l 25 l 10 l 0 40 0 60 2 00 2 00 0 15 0 20 1 00 1 00 2 25 710100 1 ch 10 5...

Page 28: ...10 l 5 l 0 50 0 70 2 50 4 00 0 20 0 30 1 50 2 50 2 20 710220 8 ch 25 250 l 250 l 125 l 25 l 0 40 0 60 1 50 0 15 0 20 1 00 2 10 710800 8 ch 50 1200 l 1200 l 600 l 120 l 50 l 0 50 1 00 3 00 8 00 0 15 0...

Page 29: ...atenttihakemuksilla Biohit Proline on Biohit Oyj n rekister ity tavaramerkki P grund av kontinuerling utveckling f rbeh ller sig Biohit r tten till f r ndringar av specifikationer utan separat anm lan...

Reviews: