9
Option 1: Power cord with transformer
This is recommended if AC power is available at the counter location.
•
Use only the provided AC adapter, and
confirm local outlet style before
ordering:
•
US: Nema 1-15
•
Europe: CEE 7
Only connect to outdoor AC outlets with a weather-proof cover comply-
ing with local electrical codes.
3 options to connect the
BG-Counter 2 to power via
the 12 V power connection
Option 2: 12 V battery (provided by user)
This is recommended if the counter is operated in a location for a few days,
before being serviced or moved. The minimum number of hours of operation
possible with a full battery can be calculated as follows:
Operation hours = Battery capacity (Ah) x 2.
For example, with a battery capacity of 60 Ah (Ampere hours), the counter and
fan can be operated for at least 60 x 2 = 120 hours = 5 full days.
Use only “deep cycle” lead-acid batteries. Normal car batteries are not desig
-
ned for continuous charge/discharge cycling, and thus would have significantly
reduced lifetime.
•
Connect battery using battery cable
Connecting the BG-Counter 2 to Power: 3 Options