6.1.2 Replacement Parts
Catalog
Number
Description
900-8933
Replacement Fuse, 8 A, 250 V, 5 x 20 mm, type T
165-9005
Power Cable
165-5058
Temperature Probe
6.2 Troubleshooting
Problem Possible
Cause
Solution
No display/lights/fan
1. No AC power.
1. Check if the PowerPac HV
is unplugged, or if there is a
problem with the AC power
source, or if the power
switch is in the “off” position.
2. Blown fuse
2. Replace the fuse. See
Section 6.1.1 for details.
Repeatedly blown fuses
Hardware failure
Contact Bio-Rad Technical
Support.
Leads from the cell are
Some leads are not
The PowerPac HV not
not long enough to fit
long enough to make
compatible with any adaptors or
the output jacks
electrical connection.
banana plugs. The PowerPac
The output terminals for the
HV can only be operated safely
PowerPac HV are
with the included connectors.
recessed 27 mm to
meet safety regulations.
Clock settings are lost
Clock battery is low.
Contact Bio-Rad Technical
Support.