4.3.1 Create, Edit, and Run an IEF Method
This section describes how to create and edit methods.
Step Procedure
Description
1.
Set up the power supply
Refer to Section 3 for instrument setup.
2.
Select the operation
At the Home screen, press the
IEF METHODS
soft
mode
key to display the IEF Method List screen.
3.
IEF Method List
Options:
• Press
NEW
to create a new method (go to step 4).
New method
•
OPEN
a stored method. Select a stored method
using the arrow scroll keys, an asterisk indicates
the selected method, and press the OPEN soft
key. Go to step 4 to edit method, otherwise go to
step 11 to start the run.
•
DELETE
a stored method. An asterisk indicates
the selected method. Select a stored method
using the arrow scroll keys and press the
DELETE
Soft key.
Existing method
Note
: If the method list is full (nine methods), a
method must be deleted before a new method
can be added to the list.
19
SELECT MODE
METHODS
MANUAL
IEF
BASIC
IEF METHOD LIST
* pH-3-10
... EMPTY 2 ...
... EMPTY 3 ...
OPEN
DELETE
NEW
METHOD: pH-3-10
S1 500V
00:05 HRS
S2 5000V
05:00 HRS
*S3 500V
00:30 HRS
METHODS DELETE
PowerPac HV – Instruction Manual