Step Procedure
Description
4.
Method name
Name a method.
• Press the up arrow key to select the method
name field.
• Use the
CE
button to clear the name field and the
alphanumeric keypad to enter a method name.
Refer to Section 2 for help using the keypad.
• Press
OK
to accept the new name and return to
the
METHOD SCREE
N with new or edited run
name.
Press
CANCEL
to abort naming the method and
return to the
METHOD SCREEN
.
5.
Method steps
Enter the Step Edit sequence.
• Press the up and down arrows to select an
existing step or select the (EMPTY) step to add a
new step.
• Press
EDIT
to enter the edit loop for the selected
step.
6.
Set the Voltage profile
• Use the
V-MAX
soft key to enter the maximum
voltage value with the keypad for the current step
• Use the
V-SLOPE
soft key to select the voltage
ramping (slow, linear or rapid).
Press
EDIT
to proceed to step 8, Edit Limits.
Use the
DELAY
soft key, available only in the
first step of a method, to program a run delay. No
voltage is applied during the duration of the step.
7.
Set time delay
• Enter the time delay and press
OK
to accept, and
go to the method steps screen (step 5) to edit
additional steps.
•
CANCEL
to return to the Volt profile screen.
20
METHOD: pH-3-10
S1 500V
00:05 HRS
S2 5000V
05:00 HRS
*S3 500V
00:30 HRS
METHODS DELETE
S1/1 VOLT PROFILE
V-SLOPE
RAPID
TIME: 01:00 HRS
V-MAX V-SLOPE DELAY
S1/1 EDIT TIME DELAY
DELAY: 00:00 HRS
CANCEL OK
PowerPac HV – Instruction Manual