Step Procedure
Description
8.
Select the time mode
Use the soft keys to select a time mode and the
alphanumeric keypad to enter the run time.
HRS
:
Run time is in units of hours.
VHOURS
:
Run time is in units of volt-hours.
UNTIMED
: Run is continuous.
Note
: Steps programmed after an untimed step are
ignored.
Press
EDIT
to return to the Method Screen.
9.
Add a new step or
To program additional steps press the down arrow
save the method
key to select the empty step and repeat Steps 5–8.
If the method will not be saved, go directly to Step 10.
Press
SAVE
and then go to Step 10.
Press
DELETE
to delete a selected step.
Press the
METHOD
soft key to return to the Method
List.
10.
Start the run
Press
to start a run.
11.
Monitor the run
The
RUN
screen is used to monitor run parameters
or to pause a run. The current step, Volts, mA and
Watts values, elapsed time versus the programmed
time, and total elapsed time are displayed on the
RUN
screen. The method step is also indicated.
This run is executing Step 1 of 2.
The run may be paused by pressing the
PAUSE
soft key or the
key.
To edit the run parameters during
PAUSE
see
Section 4.5. Edits made during the run may be
saved, if the
SAVE
soft key is pressed at the end of
the run. Otherwise the method will remain as
originally created.
To terminate the run, press the
key.
12.
The Run Completed or Run Terminated screen is
displayed at the end of a run. Press
RERUN
to
repeat the run or
EXIT
to display the method step
list.
17
500 mA
400 W
S1/1 EDIT TIME
TIME: UNTIMED
HRS VHOURS UNTIMED