PARAMETER DETAILS
1.2 AFES (AUTO FALSE ECHO SUPPRESSION) WIZARD
Note: Before using AFES, configure the device via the Quick Start wizard.
If you have a vessel with known obstructions, we recommend using AFES to prevent false echo detection.
This feature can also be used if the unit displays a false high level, or the reading is fluctuating between the correct level
and a false high level.
1. Make sure the material level is below all known obstructions.
2. Navigate to Level Meter > Diagnostics (3.) > Echo Profile (3.2.) and press RIGHT arrow to request a profile.
3. Determine a range that includes the false echo but not the true echo: in the example, 3.3 m.
4. Open the AFES wizard and press DOWN arrow to continue.
5. Press RIGHT arrow to edit Auto False Echo Suppression Range.
6. Enter the new range value and press RIGHT arrow to transfer it.
7. Press DOWN arrow to initiate learn. A transition screen appears, followed by
the message ‘Wizard Complete’.
8. Press DOWN arrow (Finish) to save AFES parameter changes and return to
Program menu.
9. Navigate to Echo Profile (3.2) and press RIGHT arrow to request a profile. The
false echo is now screeened out and the true echo selected.
10. Press LEFT arrow key to return to Measurement Mode.
False Echo
True Echo
AFES
Range
1.3 COPY PARAMETERS TO DISPLAY
Transfer parameter settings from a device to the local display interface.
1. Press RIGHT arrow to edit.
2. Press DOWN arrow to select START and RIGHT arrow key to begin the transfer.
3. PARAM UPLOAD will be displayed, then the device returns to Measurement Mode.
1.4 COPY PARAMETERS FROM DISPLAY
Transfer parameter settings from the local display interface to a device.
1. Press RIGHT arrow to edit.
2. Press DOWN arrow to select START and RIGHT arrow key to begin the transfer.
3. PARAM DOWNLOAD will be displayed, then the device returns to Measurement Mode.