11
www.bindicator.com
MPR180416 Rev. B
Cable Gland: 1/2” NPT
Cable Shield
Instrument Shield
Connection
HAZARDOUS CONNECTION
•
SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS 1, DIVISION 2
•
DO NOT DISCONNECT EQUIPMENT UNLESS POWER HAS BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-
HAZARDOUS
•
INSTALLATION SHALL BE DONE IN ACCORDANCE WITH THE N.E.C. AND C.E.C. PART ONE
•
USE APPROVED WATER TIGHT CONDUIT FITTINGS FOR OUTDOOR APPLICATIONS
IC: 21136- AIR RADAR
FCC ID: 2AHEN-AIR RADAR
THIS DEVICE COMPLIES WITH
PART 15 OF THE FCC RULES.
OPERATION IS SUBJECT TO
THE FOLLOWING TWO
CONDITIONS:
1) THIS DEVICE MAY NOT
CAUSE HARMFUL
INTERFERENCE AND
2) THIS DEVICE MUST ACCEPT
ANY INTERFERENCE
RECEIVED, INCLUDING
INTERFERENCE THAT MAY
CAUSE UNDESIRED OPERATION
VENTURE MEASUREMENT, SPARTANBURG, SC USA
24
NOM., 30 V
4 - 20 mA
OUTPUT: HART
AMB. TEMP. : – 40°C to 80°C
INPUT :
V
MAX.,
AIR RADAR
SERIAL NO.: GYZ / A1034567
ENCLOSURE: NEMA / TYPE 4X, 6; IP68
CLASS II, DIV. 1, GR. E, F, G
CLASS III, T4
CLASS I, DIV. 2
GR. A, B, C, D
T4
REFER TO INSTALLATION
DWG. A5E37393148
POTENTIAL ELECTROSTATIC
CHARGING HAZARD, DO NOT
CLEAN WITH DRY CLOTH
DO NOT INSTALL WHERE
BUILD-UP OF
CHARGE IS LIKELY
DE-ENERGIZE BEFORE
REMOVING COVER
USE SUITABLY RATED CABLE
Assembled in Canada with domestic and imported parts