CB-S / CB-S-UL (E7) 06/2020
page 62/119
15.4.7 Alarms referring to temperature sensor failures
The alarms occur immediately.
•
Alarm messages corresponding to the individual alarm cause, “collective alarm” icon
•
Audible alarm: buzzer (intermittent sound)
•
Switching the zero-voltage relay alarm output
•
A sensor fault alarm display takes priority over all other operational displays and alarm signals on the
controller.
Actions:
•
Turn off the chamber.
•
If necessary, clean and disinfect the chamber. Automatic sterilization is not possible with this fault.
•
Please contact BINDER Service.
Failure of inner temperature sensor
•
Actual temperature value display shows “ – – – – ”
•
Alarm message “Inner temp. sensor”
Temperature control continues using the safety controller temperature sensor.
Failure of safety controller temperature sensor
•
Alarm messages “Safety controller” and “Safety control sensor” alternating, “collective alarm” icon
All heating elements turn off.
Failure of inner temperature sensor and safety controller temperature sensor
•
Actual temperature value display shows “ – – – – ”
•
Alarm messages “Inner temp. sensor”, “Safety controller”, and “Safety control sensor” alternating “col-
lective alarm” icon
All heating elements turn off.
Failure of temperature sensor for door heating
•
Alarm message “Door heating sensor”, “collective alarm” icon
The door heating turns off.
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...